本研究分析使用与不使用3D打印个性化模型(3D PPM)行CT引导下三叉神经节射频治疗的参数。结果显示两组疼痛缓解程度相当,术后未观察面部麻木,但使用3D PPM组辐射时间、剂量面积乘积和空气比释动能参考点均显著低于未使用组,穿刺疼痛严重程度分布亦不同。

要点

  • 使用3D PPM可控制进针,减少患者和医务人员的辐射剂量。
  • 使用3D PPM可减轻卵圆孔区域穿刺时的疼痛。
  • 两组术后均未观察面部麻木,无严重围手术期并发症。

核心结论

  • 使用3D PPM允许控制进针,减少对患者和医务人员的辐射剂量,减轻穿刺针插入卵圆孔区域的疼痛,并最大限度降低术后并发症风险。

研究设计与样本

  • 分析使用与不使用3D PPM的CT引导下三叉神经节射频治疗的实施情况。

主要发现

  • 两组疼痛严重程度减轻相当,术后未观察面部麻木。
  • 辐射时间:组1为181.67 ± 2.99秒,组2为310.50 ± 18.46秒(p < 0.001)。
  • 剂量面积乘积:组1为950.97 ± 115.41 µGy*m2,组2为1545.48 ± 135.04 µGy*m2(p < 0.005)。
  • 空气比释动能参考点:组1为114.53 ± 16.81 mGy,组2为190.88 ± 17.48 mGy(p < 0.005)。
  • 穿刺针插入时疼痛严重程度:组1轻度62.5%、中度37.5%、重度0%;组2轻度25%、中度41.6%、重度33.3%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The purpose of this study was to present a newly designed 3D-printed personalized model (3D PPM) of a radiofrequency needle guide with a maxillary fixation for gasserian ganglion (GG) puncture. METHODS: Implementation of 3D CT-guided radiofrequency therapy of the GG with and without use of 3D PPM was analyzed. The following parameters were assessed: radiation time, dose area product, air kerma reference point, pain severity during the puncture needle insertion, prosopalgia regression degree (according to visual analog scale) and the severity of facial numbness (according to the Barrow Neurological Institute scale) in the early postoperative period, and postpuncture complications. RESULTS: Pain severity reduction was equivalent in both groups, and postoperative facial numbness was not observed. A statistically significant difference in radiation exposure parameters was revealed: radiation time was 181.67 ± 2.99 and 310.50 ± 18.46 seconds (p < 0.001); dose area product was 950.97 ± 115.41 and 1545.48 ± 135.04 µGy*m2 (p < 0.005); and the air kerma reference point was 114.53 ± 16.81 and 190.88 ± 17.48 mGy (p < 0.005) in groups 1 and 2, respectively. The severity of pain during a puncture needle insertion was assessed as mild in 62.5% and 25%, moderate in 37.5% and 41.6%, and severe in 0% and 33.3% of patients in groups 1 and 2, respectively. No serious perioperative complications were observed. CONCLUSIONS: The use of 3D PPM allows for controlled needle insertion, reducing the radiation dose to the patient and medical staff, reducing pain during a puncture needle insertion into the area of the foramen ovale, and minimizing the risk of postoperative complications.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Sufianov AA, Garifullina NA, Zyryanov AN, et al. Designing and clinical application of a 3D-printed personalized model of a radiofrequency needle guide with a maxillary fixator for puncture of the gasserian ganglion for trigeminal neuralgia treatment. J Neurosurg. 2025. PMID: 39823582. doi:10.3171/2024.8.JNS24196