文献速递功能神经外科

结合第三脑室解剖优化丘脑中央中核脑深部电刺激的间接靶向

Optimizing indirect targeting of the centromedian nucleus for deep brain stimulation by incorporating third ventricular anatomy.
PubMed 文献 · J Neurosurg · 2024-08-09

研究在100名健康成人MRI上模拟丘脑中央中核(CM)标准间接靶向并映射至共同脑模板,用 probable electrode location(PEL)掩模评估电极触点位置误差;标准间接坐标平均欧氏误差4.40±1.05 mm,改进为第三脑室后外侧角外侧4.5 mm、后连合水平后平均欧氏误差降至1.24±0.5 mm,PEL掩模落入CM比例由52%升至84%。

要点

  • 标准间接坐标平均欧氏误差4.40±1.05 mm,PEL掩模与CM重叠52%、与丘脑腹后内侧核重叠65%。
  • 第三脑室宽度变异是误差的主要来源(r=-0.71)。
  • 改进坐标为第三脑室后外侧角外侧4.5 mm、后连合水平,平均欧氏误差1.24±0.5 mm,84%的PEL掩模位于CM内。

核心结论

  • 标准间接坐标不能实现CM的最优靶向,多数触点偏向外侧位于感觉丘脑;作者提出的间接方法可能提高CM-DBS的准确性和可及性,同时减少副作用。

研究设计与样本

  • 在100名健康成人的100幅MR图像上进行CM间接靶向并将坐标映射至共同脑模板,开发 probable electrode location(PEL)掩模以估计电极触点位置,并用20例全面性癫痫患者的MR图像验证。

主要发现

  • 标准间接坐标下平均欧氏误差为4.40±1.05 mm,x、y、z误差分别为4.19±0.97 mm、0.73±0.65 mm、0.66±0.69 mm。
  • PEL掩模与CM重叠52%,与丘脑腹后内侧核重叠65%。
  • 第三脑室宽度变异是这些误差的主要来源(r=-0.71)。
  • 改进坐标下平均欧氏误差降至1.24±0.5 mm,84%的PEL掩模位于CM内。

关键数值与推荐

  • 改进的间接坐标为第三脑室后外侧角外侧4.5 mm、后连合水平。
  • 标准间接坐标平均欧氏误差4.40±1.05 mm,改进后为1.24±0.5 mm。
  • 标准坐标下PEL掩模与CM重叠52%,改进后84%位于CM内。

临床怎么用

  • 作者认为所提出的间接方法可能提高CM-DBS的准确性和可及性,同时减少副作用;原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Deep brain stimulation (DBS) of the centromedian nucleus (CM) is used to treat diverse brain diseases including epilepsy, Tourette syndrome, and disorders of consciousness. However, the CM is challenging to visualize on routine MRI. Many surgeons use an indirect targeting method based on established stereotactic coordinates. The authors aimed to quantify how often a DBS electrode's contacts were positioned within the CM using this approach, and to identify alternative indirect coordinates that are more accurate. METHODS: Indirect targeting of the CM was performed on 100 MR images obtained in healthy adults, and the resulting coordinates were warped to a common brain template. To estimate positions of DBS contacts along this trajectory, the authors developed a probable electrode location (PEL) mask, modeled on trajectory angles obtained from prior clinical cases. Euclidean and x, y, and z distances between the centroids of the PEL mask and an atlas-based CM mask were measured and defined as error. The percentage of overlaps between the PEL mask and nearby thalamic nuclei was determined. To assess the clinical utility of this methodology, the analysis was validated using 20 MR images obtained in patients with generalized epilepsy, a common clinical indication for CM-DBS. RESULTS: Using standard indirect coordinates, the authors found the average Euclidean error to be 4.40 ± 1.05 mm, and the x, y, and z errors were 4.19 ± 0.97 mm, 0.73 ± 0.65 mm, and 0.66 ± 0.69 mm, respectively. The PEL mask overlap was 52% with the CM and 65% with the ventral posteromedial nucleus. Variation in third ventricular width was the dominant contributor to these errors (r = -0.71). To overcome this variation, the authors developed alternative indirect coordinates: 4.5 mm lateral to the posterolateral corner of the third ventricle at the level of the posterior commissure. With this refinement, the average Euclidean error was reduced to 1.24 ± 0.5 mm, with 84% of the PEL mask within the CM. CONCLUSIONS: The unavailability of advanced MRI for direct targeting limits access to CM-DBS in resource-constrained neurosurgical programs. Standard indirect coordinates do not provide optimal targeting of the CM, with most contacts laterally placed in the sensory thalamus. The proposed indirect approach may therefore increase the accuracy and availability of CM-DBS, while reducing side effects.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Singh H, Stamm M, Warren AEL, et al. Optimizing indirect targeting of the centromedian nucleus for deep brain stimulation by incorporating third ventricular anatomy. J Neurosurg. 2024. PMID: 39126726. doi:10.3171/2024.4.JNS24286

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。