术中三维透视可准确预测脑深部电刺激手术的最终电极位置
本回顾性研究纳入64例(124根电极)在本机构接受脑深部电刺激(DBS)手术的患者,比较术中三维透视(3DF)与术后容积CT确定的电极尖端位置偏差、气颅影响及辐射暴露。结果显示两种方法定位高度相关(p=0.91;p<0.0001),3DF辐射暴露显著低于CT(0.36±0.03 vs 2.08±0.05 mSv;p<0.0001)。作者认为术中3DF在确定最终DBS电极位置方面与CT相当,是替代术后CT的有效工具。
要点
- 术中3DF与术后CT确定的电极尖端位置高度相关(p=0.91;p<0.0001)。
- 3DF辐射暴露显著低于CT(0.36±0.03 vs 2.08±0.05 mSv;p<0.0001)。
- 3DF组气颅体积大于CT组(6.89±1.76 vs 5.18±1.37 mm3,p<0.001),但与电极位置差异无相关性(p=0.17;p=0.06)。
核心结论
- 术中3DF在确定DBS最终电极位置方面与CT相当,且辐射暴露更低、费用更少、速度更快、避免患者转运出手术室,作者认为其是替代术后CT的有效工具。
研究设计与样本
- 回顾性研究,纳入64例在本机构接受DBS手术的患者(共124根电极),采集电极植入后术中3DF和术后容积CT图像。
主要发现
- 3DF与CT估计的电极尖端差异为0.85±0.03 mm,两种方法到MCP的距离无显著差异(p=0.11),但高度相关(p=0.91;p<0.0001)。
- 3DF组气颅体积大于CT组(6.89±1.76 vs 5.18±1.37 mm3,p<0.001),但气颅与两种技术测量的电极位置差异无相关性(p=0.17;p=0.06)。
- 3DF辐射暴露显著低于CT(0.36±0.03 vs 2.08±0.05 mSv;p<0.0001)。
临床怎么用
- 作者认为术中3DF可替代术后CT用于确定DBS最终电极位置。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: In the absence of an intraoperative CT or MRI setup, post-implantation confirmation of electrode position in deep brain stimulation (DBS) requires patient transportation to the radiology unit, prolonging surgery time. This project aims to validate intraoperative 3D fluoroscopy (3DF), a widely available tool in Neurosurgical units, as a method to determine final electrode position. METHODS: We performed a retrospective study including 64 patients (124 electrodes) who underwent DBS at our institution. Intraoperative 3DF after electrode implantation and postoperative volumetric CT were acquired. The Euclidean coordinates of the electrode tip displayed in both imaging modalities were determined and inter-method deviations were assessed. Pneumocephalus was quantified and its potential impact in determining the electrode position analyzed. Finally, 3DF and CT-imposed exposure to radiation was compared. RESULTS: The difference between the electrode tip estimated by 3DF and CT was 0.85 ± 0.03 mm, and not significantly different (p = 0.11 for the distance to MCP assessed by both methods), but was, instead, highly correlated (p = 0.91; p < 0.0001). Even though pneumocephalus was larger in 3DF (6.89 ± 1.76 vs 5.18 ± 1.37 mm3 in the CT group, p < 0.001), it was not correlated with the difference in electrode position measured by both techniques (p = 0.17; p = 0.06). Radiation exposure from 3DF is significantly lower than CT (0.36 ± 0.03 vs 2.08 ± 0.05 mSv; p < 0.0001). CONCLUSIONS: Intraoperative 3DF is comparable to CT in determining the final DBS electrode position. Being a method with fewer radiation exposure, less expensive, faster and that avoids patient transportation outside the operation room, it is a valid tool to replace postoperative CT.
引用
- Neto-Fernandes P, Chamadoira C, Silva C, et al. Intraoperative 3D fluoroscopy accurately predicts final electrode position in deep brain stimulation surgery. Acta Neurochir (Wien). 2024. PMID: 39107666. doi:10.1007/s00701-024-06214-8