PubMed 文献 · J Neurooncol · 2024-08-28

该研究在常规脑肿瘤手术中采集218个原位拉曼光谱与近红外自发荧光数据,并以375个离体光谱训练组织分类模型,发现脑肿瘤自发荧光强度低于肿瘤边界与正常脑组织,原位自发荧光与拉曼光谱分类正确率分别为0.83与0.84,与术者判断及组织病理比较结果相似,拉曼光谱确认了检测脑肿瘤的能力,自发荧光成为术中肿瘤勾画的竞争性方法。

要点

  • 脑肿瘤自发荧光强度低于肿瘤边界与正常脑组织。
  • 原位自发荧光与拉曼光谱分类正确率分别为0.83与0.84。
  • 与组织病理比较,自发荧光与拉曼光谱正确率分别为0.80与0.83。

核心结论

  • 光谱技术成功整合入现有神经外科工作流程,原位光谱数据可基于离体数据分类。
  • 拉曼光谱确认了检测脑肿瘤的能力,自发荧光成为术中肿瘤勾画的竞争性方法。

研究设计与样本

  • 在常规脑肿瘤手术中于实性肿瘤内及肿瘤边界采集原位拉曼光谱与自发荧光,共218个光谱;胶质瘤WHO II-III级n=6、胶质母细胞瘤n=10、转移瘤n=10、脑膜瘤n=3。
  • 肿瘤识别组织分类模型基于离体数据训练,共375个光谱;胶质瘤/胶质母细胞瘤患者n=20、转移瘤n=11、脑膜瘤n=13、癫痫海马n=4。

主要发现

  • 原位与离体数据均显示脑肿瘤自发荧光强度低于肿瘤边界区域与正常脑组织。
  • 自发荧光强度与1437 cm-1处脂质拉曼谱带强度呈正相关,与1260 cm-1处蛋白谱带强度呈负相关。
  • 原位自发荧光与拉曼光谱数据集分类与术者组织类型评估相符,正确率分别为0.83与0.84。
  • 与选定测量位置切除组织活检的组织病理比较,正确率相似(自发荧光:0.80,拉曼光谱:0.83)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Raman spectroscopy (RS) is a promising method for brain tumor detection. Near-infrared autofluorescence (AF) acquired during RS provides additional useful information for tumor identification and was investigated in comparison with RS for delineating brain tumors in situ. METHODS: Raman spectra were acquired together with AF in situ within the solid tumor and at the tumor border during routine brain tumor surgeries (218 spectra; glioma WHO II-III, n = 6; GBM, n = 10; metastases, n = 10; meningioma, n = 3). Tissue classification for tumor identification in situ was trained on ex vivo data (375 spectra; glioma/GBM patients, n = 20; metastases, n = 11; meningioma, n = 13; and epileptic hippocampi, n = 4). RESULTS: Both in situ and ex vivo data showed that AF intensity in brain tumors was lower than that in border regions and normal brain tissue. Moreover, a positive correlation was observed between the AF intensity and the intensity of the Raman band corresponding to lipids at 1437 cm- 1, while a negative correlation was found with the intensity of the protein band at 1260 cm- 1. The classification of in situ AF and RS datasets matched the surgeon's evaluation of tissue type, with correct rates of 0.83 and 0.84, respectively. Similar correct rates were achieved in comparison to histopathology of tissue biopsies resected in selected measurement positions (AF: 0.80, RS: 0.83). CONCLUSIONS: Spectroscopy was successfully integrated into existing neurosurgical workflows, and in situ spectroscopic data could be classified based on ex vivo data. RS confirmed its ability to detect brain tumors, while AF emerged as a competitive method for intraoperative tumor delineation.