陶瓷动脉瘤夹配钛弹簧的设计与评估

Ceramic Aneurysm Clip with Titanium Spring.
PubMed 文献 · World Neurosurg · 2024-06-26

研究设计并评估了氧化锆体部配钛弹簧的陶瓷动脉瘤夹,测量其30次开合后的闭合力,并在3特斯拉MRI下比较其与Yasargil钛合金夹的伪影。陶瓷夹初始闭合力1.70 N,30次循环后降至1.22 N,保留72%;MRI测量差异显著低于钛合金夹,L2点差异为3%。

要点

  • 陶瓷夹初始闭合力1.70 N,30次循环后1.22 N,保留72%。
  • 3特斯拉MRI下陶瓷夹测量差异显著低于Yasargil钛合金夹。
  • 陶瓷夹在高分辨率T1加权图像上图像更清晰。

核心结论

  • 陶瓷夹配钛弹簧显示出满意的闭合力和更优的MRI兼容性,有望改善手术应用与术后评估。

研究设计与样本

  • 使用Creo Parametric三维CAD软件设计陶瓷动脉瘤夹,体部为氧化锆,弹簧为钛。
  • 闭合力用测力计系统测量,陶瓷夹与Mizuho永久夹对比,进行30次开合循环。
  • MRI伪影评估使用3特斯拉扫描仪,比较陶瓷夹与Yasargil夹。

主要发现

  • 直型陶瓷夹初始闭合力1.70 N,30次循环后降至1.22 N,保留初始力的72%。
  • MRI中陶瓷夹测量差异显著低于钛合金Yasargil夹,尤其在高分辨率T1加权图像。
  • 测量点L2处陶瓷夹差异最低,为3%。
  • 陶瓷夹图像较钛合金夹更清晰,尤其在夹端部。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To design and evaluate ceramic aneurysm clips with integrated titanium springs, focusing on ergonomic application and precision in neurosurgical procedures. METHODS: The clip design was executed with precision using Creo Parametric 3D CAD software. It comprises a zirconia body and a titanium spring for durability and consistent tension and features a four-coil hairpin titanium spring for enhanced closing force and a ball-type head for versatile maneuverability during surgery. To assess durability, closing forces were rigorously measured using a force gauge system, comparing the ceramic clip with the standard Mizuho permanent clip over 30 open-close cycles. For the assessment of magnetic resonance (MR) artifacts, both the ceramic and Yasargil clips were evaluated using a 3 Tesla (T) MRI scanner, employing specific imaging sequences. RESULTS: The straight type ceramic clip's initial closing force was 1.70 N, dropping to 1.22 N after 30 cycles, indicating a retention of 72% of its initial force. In MRI, the ceramic clip displayed significantly lower measurement discrepancies compared to the titanium alloy Yasargil clip, particularly in high-resolution T1-weighted images. The lowest variance was at measurement point L2, where the ceramic clip showed a 3% discrepancy. Furthermore, the ceramic clip yielded clearer images than the titanium alloy clip, particularly at the clip's end. CONCLUSIONS: Ceramic clips with titanium springs demonstrated satisfactory closing force and superior MRI compatibility, promising enhancements in surgical application and postoperative assessment.

来源:World Neurosurg,查看原文。版权归原作者所有。