开发可调式患者特异性刚性导引器以提高脑室外引流管置入精度

Development of an adjustable patient-specific rigid guide to improve the accuracy of external ventricular catheter placement.
PubMed 文献 · J Neurosurg · 2024-05-10

作者设计了一种患者特异性脑室外引流(EVD)导引器,由三脚架底座和多个不同角度插入件组成,可在多个旋转位置锁定以提供多种置入角度。在模拟正常与变形脑室解剖的3D打印颅骨模型上测试,经冠状位和矢状位CT确认导管均准确置入脑室,包括垂直轨迹(如Ghajar导引器)不足的病例。作者认为该导引器可实现一致且准确的EVD置入,需进一步试验与徒手方法比较。

要点

  • 徒手EVD置入报告的不准确率为12.3%-44.9%,尤其在脑室解剖改变时。
  • 患者特异性EVD导引器由三脚架底座和不同角度插入件组成,可多旋转位置锁定。
  • 在模拟正常及变形脑室的颅骨模型上,导引器均实现准确置入,需进一步试验与徒手方法比较。

核心结论

  • 患者特异性EVD导引器在模拟正常和变形脑室解剖的颅骨模型中可实现一致且准确的EVD置入。
  • 需要进一步试验将该装置与徒手方法进行比较。

研究设计与样本

  • 作者设计患者特异性EVD导引器,并在3D打印颅骨模型(含模拟正常与变形脑室解剖的明胶脑模拟物)上进行测试。
  • 使用模型低分辨率CT扫描计算相对于标准垂直入路的置入角度,置入后以重复CT扫描评估准确性。

主要发现

  • 在正常脑室解剖以及异常移位脑室的试验中,正确使用患者特异性EVD导引器均实现导管准确置入脑室,经冠状位和矢状位CT图像确认。
  • 准确置入包括垂直轨迹(如Ghajar导引器)不足的病例。

临床怎么用

  • 作者认为该导引器可实现一致且准确的EVD置入,需进一步试验与徒手方法比较。

局限性

  • 需要进一步试验将该装置与徒手方法进行比较。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The most common method for external ventricular drain (EVD) placement is the freehand approach, which has reported inaccuracy rates of 12.3%-44.9%, especially in the case of altered ventricular anatomy. Current assistive devices require added time or equipment or do not account for shifted ventricles. To improve the accuracy of emergent EVD placement in the setting of altered ventricular anatomy, the authors designed a patient-specific EVD (PS-EVD) guide. METHODS: The PS-EVD guide has a tripod base and a series of differently angled inserts that lock in place at multiple rotational positions, allowing for numerous insertion angles. For testing, the authors designed a 3D-printed phantom skull with a gelatin brain analog containing ventricles simulating normal and altered ventricular anatomy. Low-resolution CT scans of the phantom were used to calculate the insertion angle in relation to the standard perpendicular entry. The corresponding insert at the correct rotational position within the base unit was positioned over the entry point on the phantom, and the catheter was inserted. Accuracy was evaluated with repeat CT scans. RESULTS: With normal ventricular anatomy, as well as abnormally shifted ventricles, proper use of the PS-EVD guide led to accurate catheter insertion into the ventricle in trials, as confirmed on coronal and sagittal CT images, including cases in which a perpendicular trajectory, such as with the Ghajar guide, was insufficient. CONCLUSIONS: The PS-EVD guide allows consistent and accurate EVD placement in phantom skulls with both normal and altered ventricular anatomy. Further trials comparing this device to the freehand approach are required.