分流治疗脑积水患者脑脊液流量的定量无创测量
本研究用快速相位对比MRI(PC-MRI)序列,先以流量体模校准(0至24 mL/hr),再对21例连续儿童分流患者扫描,测得脑脊液流量6至54 mL/hr,并与外置远端分流患者的床边集液瓶收集量比较显示良好一致,结论认为该序列可准确可靠地临床测量分流患者的脑脊液流量。
要点
- PC-MRI序列可在96秒内完成,体模校准流量范围0至24 mL/hr。
- 21例连续儿童分流患者测得脑脊液流量6至54 mL/hr。
- 外置远端分流患者中PC-MRI流量与集液瓶收集量比较显示良好一致。
核心结论
- PC-MRI序列为分流治疗患者提供了准确可靠的脑脊液流量临床测量。
- 该工作为将PC-MRI立即加入临床无创评估分流功能及未来脑脊液生理研究提供了必要框架。
研究设计与样本
- 研究使用快速(96秒)PC-MRI序列,先用已知流量范围0至24 mL/hr的流量体模校准。
- 体模校准后,对21例连续儿童患者进行PC-MRI序列扫描。
- 在5例患者中采集多次连续近端和远端测量,以检验分流系统不同部位的测量误差并确定患者内脑脊液流量变异性。
- 研究纳入首例PC-MRI测量脑脊液分流流量的在体验证,比较相位对比测得的流量与外置远端分流患者收集集液瓶中的脑脊液累积量。
主要发现
- PC-MRI序列成功测量了21例连续儿童患者的脑脊液流量,范围为6至54 mL/hr。
- 在一例外置远端分流患者中,PC-MRI流量测量与床边集液瓶收集的脑脊液体积比较显示良好一致。
- 与近端导管采集的PC-MRI测量相比,分流远端部分显示出更低的测量误差。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Standard MRI protocols lack a quantitative sequence that can be used to evaluate shunt-treated patients with a history of hydrocephalus. The objective of this study was to investigate the use of phase-contrast MRI (PC-MRI), a quantitative MR sequence, to measure CSF flow through the shunt and demonstrate PC-MRI as a useful adjunct in the clinical monitoring of shunt-treated patients. METHODS: The rapid (96 seconds) PC-MRI sequence was calibrated using a flow phantom with known flow rates ranging from 0 to 24 mL/hr. Following phantom calibration, 21 patients were scanned with the PC-MRI sequence. Multiple, successive proximal and distal measurements were gathered in 5 patients to test for measurement error in different portions of the shunt system and to determine intrapatient CSF flow variability. The study also includes the first in vivo validations of PC-MRI for CSF shunt flow by comparing phase-contrast-measured flow rate with CSF accumulation in a collection burette obtained in patients with externalized distal shunts. RESULTS: The PC-MRI sequence successfully measured CSF flow rates ranging from 6 to 54 mL/hr in 21 consecutive pediatric patients. Comparison of PC-MRI flow measurement and CSF volume collected in a bedside burette showed good agreement in a patient with an externalized distal shunt. Notably, the distal portion of the shunt demonstrated lower measurement error when compared with PC-MRI measurements acquired in the proximal catheter. CONCLUSIONS: The PC-MRI sequence provided accurate and reliable clinical measurements of CSF flow in shunt-treated patients. This work provides the necessary framework to include PC-MRI as an immediate addition to the clinical setting in the noninvasive evaluation of shunt function and in future clinical investigations of CSF physiology.
引用
- Ha JH, Borzage MT, Vanstrum EB, et al. Quantitative noninvasive measurement of cerebrospinal fluid flow in shunted hydrocephalus. J Neurosurg. 2023. PMID: 38564811. doi:10.3171/2023.7.JNS231326
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。