文献速递脊柱脊髓外科功能神经外科

脊髓刺激器桨状电极错位或移位的翻修与更换

Spinal Cord Stimulator Paddle Lead Revision and Replacement for Misplaced or Displaced Electrodes.
PubMed 文献 · World Neurosurg · 2024-03-30

该研究回顾了2021至2023年间16例因桨状电极错位而接受胸段脊髓刺激器翻修与更换的患者,电极错位表现为偏外侧(12例)、偏高(2例)或未完全位于硬膜外(2例)。所有患者均需跳跃式椎板切除或开窗,无并发症发生,随访18.4±7.3个月,疼痛强度从术前7.9±1.5降至末次随访3.6±1.7(P<0.001)。作者认为翻修更换是可行且持久的策略。

要点

  • 16例胸段桨状电极错位患者接受翻修更换,错位类型以偏外侧为主(12例)。
  • 所有患者均需跳跃式椎板切除或开窗,平均手术时间126.1±26.9分钟,无并发症。
  • 疼痛强度由术前7.9±1.5降至末次随访3.6±1.7(P<0.001),除1例外均继续使用设备。

核心结论

  • 作者认为,对错位的桨状脊髓刺激电极进行翻修与更换是可行且持久的翻修策略,即使对于长期植入并伴有广泛瘢痕的病例也是如此。

研究设计与样本

  • 回顾性病例系列,纳入2021年至2023年间因桨状电极错位接受胸段脊髓刺激器翻修与更换的16例患者,回顾病历收集人口学、手术细节及并发症发生情况。

主要发现

  • 16例患者平均年龄59.6±12.6岁,女性11例、男性5例;电极错位表现为偏外侧12例、偏高2例、未完全位于硬膜外2例。
  • 从初次植入到翻修手术的平均时间为44.8±47.5个月,平均手术时间126.1±26.9分钟,所有患者均需跳跃式椎板切除或开窗,未遇到并发症。
  • 平均随访时间18.4±7.3个月,平均术前疼痛强度7.9±1.5,末次随访时3.6±1.7(P<0.001);除1例外,其余患者在随访中继续使用其设备。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Spinal cord stimulators (SCSs) are commonly implanted via a laminotomy or laminectomy. Revision surgery may be necessary in instances of hardware failure or loss of efficacy. It is uncommon for leads to have been initially misplaced in a suboptimal position and revision in these cases necessitates additional dissection for appropriate repositioning. Accordingly, there is concern with a more extensive revision for a potentially higher risk of associated complications. This study aims to describe a series of patients with failed paddle SCS electrodes due to misplacement who underwent revision and replacement. METHODS: Patients who underwent SCS paddle replacement for misplaced paddles between 2021 and 2023 were identified. Medical charts were reviewed for demographic data, operative details, and incidence of complications. RESULTS: Sixteen patients underwent thoracic SCS paddle revision and replacement. The mean age was 59.6 ± 12.6 years, with 11 females and 5 males. Misplaced paddles were too lateral (n = 12), too high (n = 2), or incompletely within the epidural space (n = 2). The mean duration from initial implantation to revision surgery was 44.8 ± 47.5 months. The mean operative duration was 126.1 ± 26.9 minutes and all patients required a "skip" laminectomy or laminotomy. No complications were encountered. The mean length of follow-up was 18.4 ± 7.3 months. Mean preoperative pain intensity was 7.9 ± 1.5 and at last follow-up was 3.6 ± 1.7 (P < 0.001). All but 1 patient continued to use their device in follow-up. CONCLUSIONS: The revision and replacement of misplaced paddle SCS electrodes is a feasible and durable revision strategy, even in long-term implants with extensive scarring.