本研究纳入66例颅内肿瘤切除术后行术中CT的患者,用虚拟术中MRI软件将术前MRI与术中CT弹性融合,并与术后48小时内MRI比较。虚拟术中MRI与术后MRI在35例(53%)中一致(p=0.006),无假阴性,敏感度1、特异度0.26;低级别胶质瘤9例全部一致(p=0.003),4级胶质瘤19例敏感度1、特异度0.33(p=0.046),脑膜瘤或转移瘤无显著关联。

要点

  • 虚拟术中MRI预测肿瘤残留的敏感度为1(95%CI 0.86-1),特异度为0.26(95%CI 0.14-0.42)。
  • 低级别胶质瘤9例中虚拟术中MRI与术后MRI全部一致(p=0.003)。
  • 脑膜瘤或转移瘤中未发现显著关联;作者认为需在前瞻性更大样本中评估。

核心结论

  • 市售虚拟术中MRI可高敏感度预测肿瘤残留的有无。
  • 该方法在低级别胶质瘤中诊断准确率较高,在脑膜瘤或转移瘤中则低得多;作者认为这些发现必须在前瞻性更大样本研究中评估。

研究设计与样本

  • 病例系列研究,纳入66例颅内肿瘤切除术后行术中CT的患者,用虚拟术中MRI软件将术前MRI与术中CT弹性融合,并与术后48小时内MRI比较诊断准确率。

主要发现

  • 虚拟术中MRI与术后MRI在检测肿瘤残留方面一致35例(53%)(p=0.006)。
  • 未出现假阴性结果(即术后MRI有残留而虚拟术中MRI无残留)。
  • 虚拟术中MRI敏感度1(95%CI 0.86-1),特异度0.26(95%CI 0.14-0.42),阳性预测值0.44(95%CI 0.3-0.58),阴性预测值1(95%CI 0.72-1)。
  • 低级别胶质瘤9例中虚拟术中MRI与术后MRI全部一致,敏感度1(95%CI 0.59-1),特异度1(95%CI 0.16-1)(p=0.003)。
  • 4级胶质瘤19例中敏感度1(95%CI 0.69-1),特异度0.33(95%CI 0.08-0.7)(p=0.046);脑膜瘤或转移瘤中未发现显著关联。

局限性

  • 作者认为这些发现必须在前瞻性更大样本研究中评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Intraoperative MRI (iMRI) is the gold-standard technique for intraoperative evaluation of the extent of resection in brain tumor surgery. Unfortunately, it is currently available at only a few neurosurgical centers. A commercially available software, Virtual iMRI Cranial, provides an elastic fusion between preoperative MRI and intraoperative CT (iCT). The aim of this study was to evaluate the accuracy of this software in determining the presence of residual tumor. METHODS: Virtual iMRI was performed in patients who underwent iCT after intracranial tumor resection. The results of the software in terms of presence or absence of tumor residual were then compared with postoperative MRI performed within 48 hours after surgery to evaluate the diagnostic accuracy of virtual iMRI. RESULTS: Sixty-six patients were included in the present study. The virtual iMRI findings were concordant with the postoperative MRI data in 35 cases (53%) in the detection of tumor residual (p = 0.006). No false-negative findings (i.e., presence of residual on postoperative MRI and absence of residual on virtual iMRI) were encountered. Virtual iMRI had a sensitivity of 1 (95% CI 0.86-1), specificity of 0.26 (95% CI 0.14-0.42), positive predictive value of 0.44 (95% CI 0.3-0.58), and negative predictive value of 1 (95% CI 0.72-1). Subgroup analysis revealed that the virtual iMRI findings were concordant with postoperative MRI findings in all cases (n = 9) of lower-grade glioma (LGG) with a sensitivity of 1 (95% CI 0.59-1) and a specificity of 1 (95% CI 0.16-1) (p = 0.003); a statistically significant association was also found for grade 4 gliomas with a sensitivity of 1 (95% CI 0.69-1) and a specificity of 0.33 (95% CI 0.08-0.7) (p = 0.046) (19 patients). No significant association was found when considering meningiomas or metastases. CONCLUSIONS: The commercially available virtual iMRI can predict the presence or absence of tumor residual with high sensitivity. The diagnostic accuracy of this method was higher in LGGs and much lower for meningiomas or metastases; these findings must be evaluated in prospective studies in a larger population.