全身麻醉下单阶段脑深部电刺激手术立体定向系统的临床评估:技术说明
本文报告一种以颅骨安装的「钥匙」替代环形基架的立体定向系统,在8例患者中于全身麻醉下完成单阶段脑深部电刺激(DBS)电极植入与脉冲发生器植入,术后CT显示15根电极平均径向误差1.24±0.45 mm,无手术并发症。作者认为该系统可安全有效地完成 asleep 单阶段DBS手术并保持传统电极精度标准。
要点
- 8例患者全身麻醉下单阶段完成DBS电极与脉冲发生器植入,共植入15根电极。
- 靶点包括丘脑腹中间核4例、苍白球内侧部3例、丘脑底核1例。
- 术后CT示15根电极平均径向误差1.24±0.45 mm,无手术并发症。
核心结论
- 该立体定向系统有助于安全有效地完成全身麻醉下单阶段DBS手术,并保持传统电极精度标准。
研究设计与样本
- 技术说明,报告作者所在机构8例患者在全身麻醉下使用该颅骨安装「钥匙」式立体定向系统行单阶段DBS手术。
主要发现
- 8例患者共植入15根电极,靶点为丘脑腹中间核4例、苍白球内侧部3例、丘脑底核1例。
- 对苍白球内侧部和丘脑底核靶点进行了术中微电极记录。
- 术后CT显示15根电极平均径向误差1.24±0.45 mm,无手术并发症。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Conventional frame-based stereotactic systems have circumferential base frames, often necessitating deep brain stimulation (DBS) surgery in two stages: intracranial electrode insertion followed by surgical re-preparation and pulse generator implantation. Some patients do not tolerate awake surgery, underscoring the need for a safe alternative for asleep DBS surgery. A frame-based stereotactic system with a skull-mounted "key" in lieu of a circumferential base frame received US FDA clearance. The authors describe the system's application for single-stage, asleep DBS surgery in 8 patients at their institution and review its workflow and technical considerations. METHODS: Eight patients underwent DBS lead insertion and IPG implantation in a single surgical preparation under general anesthesia using the system. Postoperative CT imaging confirmed lead placement. RESULTS: Eight patients underwent implantation of 15 total leads targeting the ventral intermediate nucleus (4 patients), globus pallidus internus (GPi; 3 patients), and subthalamic nucleus (STN; 1 patient). Intraoperative microelectrode recording was conducted for GPi and STN targets. Postoperative CT imaging revealed a mean ± SD radial error of 1.24 ± 0.45 mm (n = 15 leads), without surgical complications. CONCLUSIONS: The stereotactic system facilitated safe and effective asleep, single-stage DBS surgery, maintaining traditional lead accuracy standards.
引用
- Scheitler KM, Rusheen AE, Yuen J, et al. Clinical evaluation of a stereotactic system for single-stage deep brain stimulation surgery under general anesthesia: technical note. J Neurosurg. 2024. PMID: 38489816. doi:10.3171/2024.1.JNS232563
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。