文献速递功能神经外科

7T与3T磁共振直接靶向在脑深部电刺激中的疗效比较研究

Enhancing outcomes in deep brain stimulation: a comparative study of direct targeting using 7T versus 3T MRI.
PubMed 文献 · J Neurosurg · 2024-02-23

该回顾性研究比较了7T与3T磁共振直接靶向在特发性震颤患者单侧丘脑脑深部电刺激中的疗效。共分析34例患者(每组17例),7T组震颤评定量表改善百分比显著高于3T组(64.9%±11.4% vs 50.9%±16.4%,p=0.004),平均刺激电流更低(2.0±0.8 mA vs 2.7±0.9 mA,p=0.01),电极位置方差更小(平均欧氏距离1.82±0.68 mm vs 2.75±0.81 mm,p=0.001)。作者认为7T磁共振直接靶向在适当工作流程下可减轻伪影和失真,达到相似准确性但更高精确度,从而降低刺激电流并改善震颤控制。

要点

  • 7T组震颤评定量表改善百分比显著高于3T组(64.9%±11.4% vs 50.9%±16.4%,p=0.004)。
  • 7T组平均刺激电流低于3T组(2.0±0.8 mA vs 2.7±0.9 mA,p=0.01)。
  • 7T组电极尖端与组质心的平均欧氏距离小于3T组(1.82±0.68 mm vs 2.75±0.81 mm,p=0.001),电极位置方差更小。

核心结论

  • 尽管存在7T磁共振伪影和失真增加的担忧,作者显示通过适当工作流程可减轻这些影响,使用7T磁共振直接靶向可改善手术结局;结果提示与3T磁共振相比,7T磁共振靶向准确性相似但精确度更高,导致刺激电流更低和震颤改善更好。未来需要研究评估7T磁共振靶向其他皮层下结构的结局。

研究设计与样本

  • 回顾性研究,纳入2021年至2023年间接受单侧丘脑脑深部电刺激治疗特发性震颤的患者,共分析34例,两个匹配队列各17例,分别使用7T磁共振和3T磁共振进行手术规划;主要终点为Fahn-Tolosa-Marin震颤评定量表评分改善百分比;采用非参数Mann-Whitney U检验比较人口学和临床数据,计算每个触点与组平均质心的平方欧氏距离并在整个队列中平均以提供电极触点位置的方差。

主要发现

  • 7T组震颤评定量表改善百分比显著高于3T组(64.9%±11.4% vs 50.9%±16.4%,p=0.004)。
  • 7T组平均刺激电流低于3T组(2.0±0.8 mA vs 2.7±0.9 mA,p=0.01)。
  • 7T组电极尖端与组质心的平均欧氏距离显著小于3T组(1.82±0.68 mm vs 2.75±0.81 mm,p=0.001),电极位置更聚集、方差更低。
  • 两组人口学信息、平均手术日期、术中靶点重新定位或不良事件无显著差异。

临床怎么用

  • 作者认为,尽管存在7T磁共振伪影和失真增加的担忧,通过适当工作流程可减轻这些影响,使用7T磁共振直接靶向可改善手术结局;结果提示7T磁共振靶向准确性相似但精确度更高,导致刺激电流更低和震颤改善更好。

局限性

  • 未来需要研究评估7T磁共振靶向其他皮层下结构的结局。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The aim of this study was to compare outcomes of direct targeting in deep brain stimulation (DBS) for essential tremor using 7T MRI versus 3T MRI. The authors hypothesized that 7T MRI direct targeting would be noninferior to 3T MRI in early tremor outcomes. METHODS: A retrospective study was conducted on patients undergoing unilateral thalamic DBS for essential tremor between 2021 and 2023. Two matched cohorts were assessed, one using 7T MRI and the other using 3T MRI for surgical planning. The primary endpoint was the percentage improvement in the Fahn-Tolosa-Marin Tremor Rating Scale (TRS) scores. Additionally, the authors assessed optimized programming settings and variance in electrode position on postoperative imaging. Demographic and clinical data were compared using the nonparametric Mann-Whitney U-test. The squared Euclidean distance of each contact from the group mean centroid was calculated and averaged across the entire cohort to provide the variance (i.e., the mean squared distance) of electrode contact position. RESULTS: A total of 34 patients were analyzed, with 17 in each cohort. There were no significant differences in demographic information or mean surgical dates between the groups. There were no differences in intraoperative target repositioning or adverse events. The 7T group had a significantly greater TRS improvement than the 3T group (64.9% ± 11.4% vs 50.9% ± 16.4%, p = 0.004). Patients in the 7T cohort also had a lower mean stimulation current compared with those in the 3T cohort (2.0 ± 0.8 mA vs 2.7 ± 0.9 mA, p = 0.01). Image evaluation revealed that although the mean electrode position was comparable between 7T and 3T, the 7T electrode positioning was more clustered, indicating a lower variance in the final electrode location. The mean Euclidean distance between the individual electrode tips and the group centroid was significantly less at 7T than at 3T (1.82 ± 0.68 mm vs 2.75 ± 0.81 mm, p = 0.001). CONCLUSIONS: Despite concerns for increased artifacts and distortions at 7T, the authors show that these effects can be mitigated with an appropriate workflow, leading to improved surgical outcomes with direct targeting using 7T MRI. Their results suggest similar accuracy but greater precision in targeting with 7T MRI compared with 3T MRI, resulting in lower stimulation currents and improved tremor reduction. Future studies are needed to assess outcomes related to 7T MRI in targeting other subcortical structures.

来源:J Neurosurg,查看原文。版权归原作者所有。

引用

  1. Middlebrooks EH, Tipton PW, Greco E, et al. Enhancing outcomes in deep brain stimulation: a comparative study of direct targeting using 7T versus 3T MRI. J Neurosurg. 2024. PMID: 38394660. doi:10.3171/2023.12.JNS232473

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。