非导航二维术中超声:在胶质瘤手术中达到高诊疗标准的一种简便手术工具
这项单中心回顾性研究纳入99例连续非选择的高级别和低级别胶质瘤患者,评估非导航二维术中超声(nn-2D-IOUS)在图像质量、残余肿瘤识别及切除范围方面的作用,以早期术后MRI为参考标准。nn-2D-IOUS在79%患者中生成良好质量图像,检测残余肿瘤灵敏度68%、特异度100%,以全切除为目标时86%患者切除超过初始肿瘤体积95%,未检出残余者的未检出肿瘤平均体积仅0.29 cm3。
要点
- nn-2D-IOUS在79%的胶质瘤患者中生成良好质量图像。
- 检测残余肿瘤灵敏度68%、特异度100%,以早期术后MRI为参考标准。
- 以全切除为目标时86%患者切除超过初始肿瘤体积95%,未检出残余者未检出肿瘤平均体积0.29 cm3。
核心结论
- 本研究支持nn-2D-IOUS在胶质瘤手术中具有重要价值,传统技术有助于改善手术结局,在资源受限环境中尤为相关。
研究设计与样本
- 单中心回顾性研究,纳入99例连续非选择的高级别和低级别胶质瘤患者,以早期术后MRI为参考标准评估nn-2D-IOUS。
- 试验注册号NCT05873946, retrospectively registered,注册日期2023年5月24日。
主要发现
- nn-2D-IOUS在79%的受评估患者中成功生成良好质量图像。
- nn-2D-IOUS检测术中残余肿瘤的灵敏度为68%,特异度为100%。
- 以全切除为手术目标时,86%的患者实现切除超过初始肿瘤体积95%。
- nn-2D-IOUS未检出残余肿瘤的患者中,未检出肿瘤组织的平均体积为0.29 cm3。
临床怎么用
- 原文结论支持nn-2D-IOUS在胶质瘤手术中的作用,并指出其尤其适用于资源受限环境,强调优化现有工具以实现高效患者照护。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: In an era characterized by rapid progression in neurosurgical technologies, traditional tools such as the non-navigated two-dimensional intraoperative ultrasound (nn-2D-IOUS) risk being overshadowed. Against this backdrop, this study endeavors to provide a comprehensive assessment of the clinical efficacy and surgical relevance of nn-2D-IOUS, specifically in the context of glioma resections. METHODS: This retrospective study undertaken at a single center evaluated 99 consecutive, non-selected patients diagnosed with both high-grade and low-grade gliomas. The primary objective was to assess the proficiency of nn-2D-IOUS in generating satisfactory image quality, identifying residual tumor tissue, and its influence on the extent of resection. To validate these results, early postoperative MRI data served as the reference standard. RESULTS: The nn-2D-IOUS exhibited a high level of effectiveness, successfully generating good quality images in 79% of the patients evaluated. With a sensitivity rate of 68% and a perfect specificity of 100%, nn-2D-IOUS unequivocally demonstrated its utility in intraoperative residual tumor detection. Notably, when total tumor removal was the surgical objective, a resection exceeding 95% of the initial tumor volume was achieved in 86% of patients. Additionally, patients in whom residual tumor was not detected by nn-2D-IOUS, the mean volume of undetected tumor tissue was remarkably minimal, averaging at 0.29 cm3. CONCLUSION: Our study supports nn-2D-IOUS's invaluable role in glioma surgery. The results highlight the utility of traditional technologies for enhanced surgical outcomes, even when compared to advanced alternatives. This is particularly relevant for resource-constrained settings and emphasizes optimizing existing tools for efficient patient care. NCT05873946 - 24/05/2023 - Retrospectively registered.
引用
- Cepeda S, García-García S, Arrese I, et al. Non-navigated 2D intraoperative ultrasound: An unsophisticated surgical tool to achieve high standards of care in glioma surgery. J Neurooncol. 2024. PMID: 38413458. doi:10.1007/s11060-024-04614-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。