文献速递神经肿瘤

荧光素染色共聚焦激光显微内镜与传统冰冻切片对颅内肿瘤术中组织病理学评估的比较

Fluorescein-stained confocal laser endomicroscopy versus conventional frozen section for intraoperative histopathological assessment of intracranial tumors.
PubMed 文献 · Neuro Oncol · 2024-01-01

这项II期非劣效性、前瞻性、多中心、非随机、超说明书临床试验比较荧光素染色术中共聚焦激光显微内镜(CLE)与常规术中冰冻切片(FS)的诊断准确性。共纳入210例患者,CLE诊断准确性0.87低于FS的0.91,差异0.04(95%CI -0.10;0.02;P=.367);CLE术中诊断中位时间3分钟,FS为27分钟,平均差异27.5分钟(标准差14.5;P<.001)。

要点

  • CLE诊断准确性0.87,低于FS的0.91,差异0.04(95%CI -0.10;0.02;P=.367)。
  • CLE术中诊断中位时间3分钟,FS为27分钟,平均差异27.5分钟(P<.001)。
  • 4例(2%)患者记录6起严重不良事件。

核心结论

  • CLE可安全、省时地完成术中组织学诊断,在所有纳入颅内病变中诊断准确性为87%。
  • 该技术实现实时组织学评估,处理时间较FS缩短10倍,可能即时影响手术策略。

研究设计与样本

  • II期非劣效性、前瞻性、多中心、非随机、超说明书临床试验(EudraCT: 2019-004512-58)。
  • 纳入18岁以上、计划择期切除的任何颅内病变患者,由3个中心在2020年11月至2022年6月共入组210例。

主要发现

  • 最常见病变为高级别胶质瘤(37.9%)、转移瘤(24.1%)和脑膜瘤(22.7%)。
  • 4例(2%)患者记录6起严重不良事件。
  • 主要终点:CLE诊断准确性0.87低于FS的0.91,差异0.04(95%CI -0.10;0.02;P=.367)。
  • CLE术中诊断中位时间3分钟,FS为27分钟,平均差异27.5分钟(标准差14.5;P<.001)。

临床怎么用

  • 作者认为CLE可安全、省时地完成术中组织学诊断,诊断准确性87%,处理时间较FS缩短10倍,可能即时影响手术策略。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The aim of this clinical trial was to compare Fluorescein-stained intraoperative confocal laser endomicroscopy (CLE) of intracranial lesions and evaluation by a neuropathologist with routine intraoperative frozen section (FS) assessment by neuropathology. METHODS: In this phase II noninferiority, prospective, multicenter, nonrandomized, off-label clinical trial (EudraCT: 2019-004512-58), patients above the age of 18 years with any intracranial lesion scheduled for elective resection were included. The diagnostic accuracies of both CLE and FS referenced with the final histopathological diagnosis were statistically compared in a noninferiority analysis, representing the primary endpoint. Secondary endpoints included the safety of the technique and time expedited for CLE and FS. RESULTS: A total of 210 patients were included by 3 participating sites between November 2020 and June 2022. Most common entities were high-grade gliomas (37.9%), metastases (24.1%), and meningiomas (22.7%). A total of 6 serious adverse events in 4 (2%) patients were recorded. For the primary endpoint, the diagnostic accuracy for CLE was inferior with 0.87 versus 0.91 for FS, resulting in a difference of 0.04 (95% confidence interval -0.10; 0.02; P = .367). The median time expedited until intraoperative diagnosis was 3 minutes for CLE and 27 minutes for FS, with a mean difference of 27.5 minutes (standard deviation 14.5; P < .001). CONCLUSIONS: CLE allowed for a safe and time-effective intraoperative histological diagnosis with a diagnostic accuracy of 87% across all intracranial entities included. The technique achieved histological assessments in real time with a 10-fold reduction of processing time compared to FS, which may invariably impact surgical strategy on the fly.

来源:Neuro Oncol,查看原文。版权归原作者所有。

引用

  1. Wagner A, Brielmaier MC, Kampf C, et al. Fluorescein-stained confocal laser endomicroscopy versus conventional frozen section for intraoperative histopathological assessment of intracranial tumors. Neuro Oncol. 2024. PMID: 38243410. doi:10.1093/neuonc/noae006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。