手术改变脑深部电刺激导线轨迹对3T MRI射频加热的影响:从体模实验到临床实施
该研究通过体模实验评估244种脑深部电刺激(DBS)导线轨迹在3T MRI下的射频加热,发现增加同心环数量并使环靠近手术钻孔可显著降低加热,临床植入推荐轨迹后射频加热降低三倍。
要点
- 体模实验中244种导线轨迹的最大温升范围为0.09°C至7.34°C。
- 增加同心环数量并使环靠近手术钻孔可大幅降低射频加热。
- 临床实施推荐轨迹后射频加热降低三倍。
核心结论
- 手术改变DBS导线颅外段轨迹可显著降低3T MRI期间的射频加热。
研究设计与样本
- 体模实验评估244种不同导线轨迹的最大温升,使用商用DBS系统与高比吸收率序列(B1+rms=2.7 µT)。
- 两位神经外科医生在患者中植入推荐修改轨迹,并与未修改轨迹的射频加热进行比较。
主要发现
- 最大温升范围为0.09°C至7.34°C。
- 增加环的数量并使环更靠近手术钻孔,尤其对侧导线,可大幅降低射频加热。
- 轨迹修改在手术中易于实施,并使射频加热降低三倍。
临床怎么用
- 作者认为在DBS导线植入时采用简单调整如钻孔附近的小同心环,可改善MRI期间的患者安全。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Radiofrequency (RF) tissue heating around deep brain stimulation (DBS) leads is a well-known safety risk during MRI, resulting in strict imaging guidelines and limited allowable protocols. The implanted lead's trajectory and orientation with respect to the MRI electric fields contribute to variations in the magnitude of RF heating across patients. Currently, there are no surgical requirements for implanting the extracranial portion of the DBS lead, resulting in substantial variations in clinical lead trajectories and consequently RF heating. Recent studies have shown that incorporating concentric loops in the extracranial lead trajectory can reduce RF heating. However, optimal positioning of the loops and the quantitative benefit of trajectory modification in terms of added safety margins during MRI remain unknown. In this study, the authors systematically evaluated the characteristics of DBS lead trajectories that minimize RF heating during 3T MRI to develop the best surgical practices for safe access to postoperative MRI, and they present the first surgical implementation of these modified trajectories. METHODS: The authors performed experiments to assess the maximum temperature increase of 244 distinct lead trajectories. They investigated the effect of the position, number, and size of the concentric loops on the skull. Experiments were performed in an anthropomorphic phantom implanted with a commercial DBS system, and RF exposure was generated by applying a high specific absorption rate sequence (B1+rms = 2.7 µT). The authors conducted test-retest experiments to assess the reliability of measurements. Additionally, they evaluated the effect of imaging landmarks and perturbations to the DBS device configuration on the efficacy of low-heating trajectories. Finally, two neurosurgeons implanted the recommended modified trajectories in patients, and the authors characterized their RF heating in comparison with unmodified trajectories. RESULTS: The maximum temperature increase ranged from 0.09°C to 7.34°C. The authors found that increasing the number of loops and positioning them closer to the surgical burr hole, particularly for the contralateral lead, substantially reduced RF heating. These trajectory modifications were easily incorporated during the surgical procedure and resulted in a threefold reduction in RF heating. CONCLUSIONS: Surgically modifying the extracranial portion of the DBS lead trajectory can substantially reduce RF heating during 3T MRI. The authors' results indicate that simple adjustments to the lead's configuration, such as small, concentric loops near the burr hole, can be readily adopted during DBS lead implantation to improve patient safety during MRI.
引用
- Vu J, Bhusal B, Rosenow JM, et al. Effect of surgical modification of deep brain stimulation lead trajectories on radiofrequency heating during MRI at 3T: from phantom experiments to clinical implementation. J Neurosurg. 2023. PMID: 37948679. doi:10.3171/2023.8.JNS23580
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。