帕金森病脑深部电刺激术并发症:单中心517例连续病例经验
该单中心研究分析2006至2021年517例帕金森病脑深部电刺激术(DBS)病例,平均随访4.68±2.86年,死亡率0.2%、永久性并发症0.6%,硬件相关并发症12.2%,最常见为伤口修整(7.2%);作者认为DBS总体安全,但需未来研究降低硬件相关并发症。
要点
- 517例帕金森病DBS病例中,死亡率0.2%,永久性并发症0.6%,颅内出血发生率2.5%。
- 癫痫发生率3.3%,受电极周围脑水肿和年龄显著影响。
- 硬件相关并发症发生率12.2%,最常见为伤口修整(7.2%)。
核心结论
- 作者认为DBS总体应被视为安全的手术,但未来需要研究以降低可能与手术技术和特定硬件设计均相关的硬件相关并发症发生率。
- 某些电极型号相关的电极周围脑水肿发生率增加同样值得未来研究。
研究设计与样本
- 单中心2006年至2021年共517例帕金森病DBS病例,平均随访4.68±2.86年。
- 采用Spearman相关系数分析颅内出血与记录轨道数的相关性,多元逻辑回归分析癫痫和颅内出血的潜在危险因素,Kaplan-Meier曲线分析硬件相关并发症的累积比例。
主要发现
- 死亡率为0.2%,永久性并发症为0.6%。
- 2.5%的病例发生颅内出血,未受记录所用轨道数影响。
- 3.3%报告癫痫,受电极周围脑水肿和年龄显著影响。
- 美敦力电极的电极周围脑水肿发生率显著高于波士顿科学电极(Χ2(1)=5.927,P=0.015)。
- 12.2%的植入报告硬件相关并发症,最常见为伤口修整(7.2%)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The number of deep brain stimulation (DBS) procedures is rapidly rising as well as the novel indications. Reporting adverse events related to surgery and to the hardware used is essential to define the risk-to-benefit ratio and develop novel strategies to improve it. OBJECTIVE: To analyze DBS complications (both procedure-related and hardware-related) and further assess potential predictive factors. METHODS: Five hundred seventeen cases of DBS for Parkinson's disease were performed between 2006 and 2021 in a single center (mean follow-up: 4.68 ± 2.86 years). Spearman's Rho coefficient was calculated to search for a correlation between the occurrence of intracerebral hemorrhage (ICH) and the number of recording tracks. Multiple logistic regression analyzed the probability of developing seizures and ICH given potential risk factors. Kaplan-Meier curves were performed to analyze the cumulative proportions of hardware-related complications. RESULTS: Mortality rate was 0.2%, while permanent morbidity 0.6%. 2.5% of cases suffered from ICH which were not influenced by the number of tracks used for recordings. 3.3% reported seizures that were significantly affected by perielectrode brain edema and age. The rate of perielectrode brain edema was significantly higher for Medtronic's leads compared to Boston Scientific's (Χ2(1)= 5.927, P= 0.015). 12.2% of implants reported Hardware-related complications, the most common of which were wound revisions (7.2%). Internal pulse generator models with smaller profiles displayed more favorable hardware-related complication survival curves compared to larger designs (X2(1)= 8.139, P= 0.004). CONCLUSION: Overall DBS has to be considered a safe procedure, but future research is needed to decrease the rate of hardware-related complications which may be related to both the surgical technique and to the specific hardware's design. The increased incidence of perielectrode brain edema associated with certain lead models may likewise deserve future investigation.
引用
- Servello D, Galbiati TF, Iess G, et al. Complications of deep brain stimulation in Parkinson's disease: a single-center experience of 517 consecutive cases. Acta Neurochir (Wien). 2023. PMID: 37773459. doi:10.1007/s00701-023-05799-w
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。