帕金森病脑深部电刺激术中丘脑底核电极植入准确性及影响因素:越南两中心结局

Accuracy of Subthalamic Nucleus Electrode Implantation in Deep Brain Stimulation Surgery for Parkinson's Disease Treatment and Affecting Factors: Outcomes at Two Centers in Vietnam.
PubMed 文献 · World Neurosurg · 2023-06-24

该回顾性分析纳入越南两家医院2014年6月至2021年7月间接受丘脑底核脑深部电刺激术的58例晚期帕金森病患者(共115根电极),评估框架立体定向技术下电极植入准确性。平均靶点误差1.94±0.73 mm、径向误差1.16±0.69 mm、角度误差2.22±4.24度;硬膜下积气体积、皮质移位和颅内电极弯曲与准确性存在统计学显著相关。作者认为该技术准确性较高、误差可接受,需更多临床试验直接比较影响因素。

要点

  • 58例晚期帕金森病患者共植入115根丘脑底核电极,平均靶点误差1.94±0.73 mm。
  • 硬膜下积气体积、皮质移位和颅内电极弯曲与电极植入准确性存在统计学显著相关。
  • 作者认为本中心所用丘脑底核电极植入技术准确性较高、误差可接受。

核心结论

  • 作者认为本中心应用的丘脑底核脑深部电刺激电极植入技术准确性较高、误差可接受。
  • 作者指出需更多临床试验直接比较影响电极植入准确性的因素。

研究设计与样本

  • 回顾性分析2014年6月至2021年7月在越南胡志明市Nguyen Tri Phuong医院和大学医学中心接受丘脑底核脑深部电刺激术的58例晚期帕金森病患者,共115根电极。
  • 所有患者均在局部麻醉下采用标准框架立体定向技术,结合微电极记录和宏刺激测试完成手术。

主要发现

  • 纳入26例女性(44.8%)和32例男性(55.2%),平均年龄60.4±8.3岁(范围40-76岁)。
  • 115根电极的平均靶点误差为1.94±0.73 mm,径向误差为1.16±0.69 mm,角度误差为2.22±4.24度。
  • 各坐标轴向量误差ΔX为-0.35±1.02 mm,ΔY为+0.99±0.82 mm,ΔZ为+0.73±0.99 mm。
  • 硬膜下积气体积、皮质移位和颅内电极弯曲与准确性之间存在统计学显著相关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Deep brain stimulation (DBS) surgery for Parkinson's Disease (PD) has become more and more popular in Vietnam. However, the accuracy of implantation and affecting factors are under investigation. The objective of this study is to evaluate the accuracy of the subthalamic nucleus (STN)-DBS electrode implantation technique for treatment PD at Nguyen Tri Phuong Hospital and University Medical Center. To investigate factors related to accuracy. METHODS: We carried out a retrospective analysis of 58 patients with advanced PD who underwent STN-DBS surgery at Nguyen Tri Phuong Hospital and University Medical Center in Ho Chi Minh City, Viet Nam between June 2014 and July 2021 (115 leads total). All patients underwent the procedure with standard frame-based techniques under local anesthesia with microelectrode recording and macrostimuation test. RESULTS: Twenty-six female (44.8%) and thirty-two male (55.2%) patients with a mean age of 60.4 ± 8.3 years old (40-76 years) were included. Of total of 115 electrodes implanted, the mean target error (ΔT), radial error (ΔR), angle error (Δθ) were 1.94 ± 0.73 mm; 1.16 ± 0.69 mm; 2.22 ± 4.24 degrees, respectively. Vector error on each coordinate axis ΔX, ΔY, ΔZ were -0.35 ± 1.02 mm, +0.99 ± 0.82 mm, +0.73 ± 0.99 mm, respectively. There was a statistically significant correlation between subdural air volume, cortical shift, intracranial electrode bending, and accuracy. CONCLUSIONS: The current STN-DBS electrode implantation technique applied in our centers was quite accurate with acceptable error. More clinical trials are necessary to directly compare affecting factors to the accuracy of electrode implantation.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Pham TA, Dao PD, Nguyen LT, et al. Accuracy of Subthalamic Nucleus Electrode Implantation in Deep Brain Stimulation Surgery for Parkinson's Disease Treatment and Affecting Factors: Outcomes at Two Centers in Vietnam. World Neurosurg. 2023. PMID: 37356489. doi:10.1016/j.wneu.2023.06.079

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。