文献速递功能神经外科

一种利用计算机断层扫描条纹伪影确定定向脑深部电刺激电极方向的新颖简单方法

A Novel and Simple Method Using Computed Tomography Streak Artifact to Determine the Orientation of Directional Deep Brain Stimulation Leads.
PubMed 文献 · Neurosurgery · 2023-05-25

该研究提出一种利用常规薄层CT扫描和商用立体定向软件,通过观察定向电极标记产生的条纹伪影来判断电极方向的方法,并在体外模型上进行了验证。方法在3家厂商的定向电极上均可靠,可简化编程流程。

要点

  • 定向电极标记在CT上产生平行于标记轴的高密度对称条纹伪影和正交的低密度暗带。
  • 条纹伪影常足以推断标记方向,否则可结合X线片排除相反方向。
  • 该方法在3家厂商的电极上均可靠,使用常规影像和现成软件即可完成。

核心结论

  • 作者提出一种在常规影像和现成软件上精确确定定向DBS电极方向的方法,该方法跨DBS厂商可靠,可简化流程并辅助有效编程。

研究设计与样本

  • 研究分析接受定向电极DBS患者的术后薄层CT和X线片,电极来自3家厂商,使用商用立体定向软件定位电极并规划覆盖电极的新轨迹。
  • 通过轨迹视图在垂直于电极的平面上定位定向标记并观察条纹伪影,随后用模型CT在3种不同方向电极的正交薄层CT图像上验证该方法。

主要发现

  • 定向标记产生反映电极方向的独特条纹伪影:平行于标记轴的高密度对称条纹伪影和正交于标记的对称低密度暗带。
  • 该伪影常足以推断标记方向,若不能则至少给出两个相反方向可能,可通过与X线片比较轻松解决。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND OBJECTIVES: Directional leads have garnered widespread use in deep brain stimulation (DBS) because of the ability to steer current and maximize the therapeutic window. Accurate identification of lead orientation is critical to effective programming. Although directional markers are visible on 2-dimensional imaging, precise orientation may be difficult to interpret. Recent studies have suggested methods of determining lead orientation, but these involve advanced intraoperative imaging and/or complex computational algorithms. Our objective is to develop a precise and reliable method of determining orientation of directional leads using conventional imaging techniques and readily available software. METHODS: We examined postoperative thin-cut computed tomography (CT) scans and x-rays of patients who underwent DBS with directional leads from 3 vendors. Using commercially available stereotactic software, we localized the leads and planned new trajectories precisely overlaying the leads visualized on CT. We used trajectory view to locate the directional marker in a plane orthogonal to the lead and inspected the streak artifact. We then validated this method with a phantom CT model by acquiring thin-cut CT images orthogonal to 3 different leads in various orientations confirmed under direct visualization. RESULTS: The directional marker creates a unique streak artifact that reflects the orientation of the directional lead. There is a hyperdense symmetric streak artifact parallel to the axis of the directional marker and a symmetric hypodense dark band orthogonal to the marker. This is often sufficient to infer the direction of the marker. If not, it at least renders 2 opposite possibilities for the direction of the marker, which can then be easily reconciled by comparison with x-ray images. CONCLUSION: We propose a method to determine orientation of directional DBS leads in a precise manner on conventional imaging and readily available software. This method is reliable across DBS vendors, and it can simplify this process and aid in effective programming.

来源:Neurosurgery,查看原文。版权归原作者所有。

引用

  1. Fayed I, Syed M, Gingold E, et al. A Novel and Simple Method Using Computed Tomography Streak Artifact to Determine the Orientation of Directional Deep Brain Stimulation Leads. Neurosurgery. 2023. PMID: 37227135. doi:10.1227/neu.0000000000002536

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。