自动化脑脊液引流在正常压力脑积水中的获益
本研究比较手动滴壶引流与LiquoGuard自动脑脊液管理系统在疑似正常压力脑积水患者延长腰椎引流中的表现。42例患者分为手动组31例与自动组11例,自动组容量性过度引流和引流不足均为0%,手动组分别为90.3%和38.7%(P<0.05);自动组过度引流症状更少、住院时间延长比例更低。
要点
- 手动组容量性过度引流90.3%(n=28),自动组0%(P<0.05)。
- 手动组过度引流症状54.8%(n=17),自动组18.2%(n=2)(P<0.05)。
- 手动组住院时间延长38.7%(n=12),自动组9%(n=1)(P<0.05)。
核心结论
- LiquoGuard装置在疑似正常压力脑积水患者的脑脊液引流中更优,可减少并发症和住院时间。
研究设计与样本
- 研究纳入接受延长腰椎引流的疑似正常压力脑积水患者,分为手动滴壶引流组与LiquoGuard自动引流组,共分析42例患者的数据。
主要发现
- 容量性过度引流见于手动组90.3%(n=28) vs 自动组0%(P<0.05)。
- 容量性引流不足见于手动组38.7%(n=12) vs 自动组0%(P<0.05)。
- 过度引流症状见于手动组54.8%(n=17) vs 自动组18.2%(n=2)(P<0.05),手动组所有出现症状者均有容量性过度引流,自动组出现症状者均无实际过度引流。
- 住院时间延长见于手动组38.7%(n=12) vs 自动组9%(n=1)(P<0.05)。
- 手动引流组过度引流症状更多,可能源于脑脊液引流波动增大,而非LiquoGuard系统的平稳引流。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The commonly used cerebrospinal fluid (CSF) drainage system remains the manual drip-chamber drain. The LiquoGuard (Möller Medical GmbH, Germany) is an automated CSF management device with dual functionality, measuring intracranial pressure and automatic pressure- or volume-led CSF drainage. There is limited research for comparison of devices, particularly in the neurosurgical field, where it has potential to reshape care. OBJECTIVE: This study aims to compare manual drip-chamber drain versus LiquoGuard system, by assessing accuracy of drainage, associated morbidity and impact on length of stay. METHOD: Inclusion criteria consisted of suspected normal pressure hydrocephalus (NPH) patients undergoing extended lumbar drainage. Patients were divided into manual drain group versus automated group. RESULTS: Data was analysed from 42 patients: 31 in the manual group versus 11 in the LiquoGuard group. Volumetric over-drainage was seen in 90.3% (n = 28) versus 0% (p < 0.05), and under-drainage in 38.7% (n = 12) versus 0% (p < 0.05), in the manual and automatic group, respectively. Symptoms of over-drainage were noted in 54.8% (n = 17) of the manual group, all of which had episodes of volumetric over-drainage, versus 18.2% (n = 2) in automated group, of which neither had actual over-drainage (p < 0.05). Higher over-drainage symptoms of manual drain is likely due to increased fluctuation of CSF drainage, instead of smooth CSF drainage seen with LiquoGuard system. An increased length of stay was seen in 38.7% (n = 12) versus 9% (n = 1) (p < 0.05) in the manual and LiquoGuard group, respectively. CONCLUSION: The LiquoGuard device is a more superior way of CSF drainage in suspected NPH patients, with reduced morbidity and length of stay.
引用
- Khawari S, Kneizeh M, Elborady M, et al. The benefits of automated CSF drainage in normal pressure hydrocephalus. Acta Neurochir (Wien). 2023. PMID: 36690867. doi:10.1007/s00701-023-05503-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。