基于Wi-Fi的光纤颅内压监测系统——一项体内研究
该研究在猪模型中植入无线光纤颅内压传感器与有线参考传感器,连续监测120分钟,每5分钟行颈静脉压迫以评估颅内压变化,分3阶段比较。结果显示两传感器在颈静脉压迫与释放时的颅内压变化具有正相关(阶段1 r=0.829,P<.001;阶段2 r=0.262,P<.001;阶段3 r=0.669,P<.001),作者认为该Wi-Fi颅内压监测系统可行,进一步适配正在开发中。
要点
- 在猪脑实质内植入无线光纤与有线参考颅内压传感器,连续监测120分钟。
- 每5分钟行颈静脉压迫,分3阶段比较两传感器颅内压变化的相关性。
- 阶段1 Spearman r=0.829(P<.001),阶段2 r=0.262(P<.001),阶段3 r=0.669(P<.001)。
核心结论
- 作者认为Wi-Fi颅内压监测系统以全面且可行的方式得到验证,使用智能手机可观察到颅内压变化具有足够的相关性,进一步适配正在开发中。
研究设计与样本
- 在猪模型中植入无线原型与有线参考两种颅内压传感器于脑实质,连续监测120分钟,每5分钟行颈静脉压迫以评估颅内压变化,实验分为3个阶段进行比较与分析。
主要发现
- 阶段1颈静脉压迫与释放期间两传感器颅内压变化一致,Spearman相关为正且强(r=0.829,P<.001)。
- 阶段2在骨孔内调换传感器后,Spearman相关为正且中等偏弱(r=0.262,P<.001)。
- 阶段3传感器复位至原骨孔后,原型表现与参考传感器相似,Spearman相关为正且中等偏强(r=0.669,P<.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Continuous invasive monitoring of intracranial pressure (ICP) is essential in neurocritical care for surveillance and management of raised ICP. Fluid-based systems and strain gauge microsensors remain the current standard. In the past few decades, several studies with wireless monitoring were developed aiming to reduce invasiveness and complications. OBJECTIVE: To describe a novel Wi-Fi fiber-optic device for continuous ICP monitoring using smartphone in a swine model. METHODS: Two ICP sensors (wireless prototype and wire-based reference) were implanted in the cerebral parenchyma of a swine model for a total of 120 minutes of continuous monitoring. Every 5 minutes, jugular veins compression was performed to evaluate ICP changes. The experimentation was divided in 3 phases for comparison and analysis. RESULTS: Phase 1 showed agreement in ICP changes for both sensors during jugular compression and releasing, with a positive and strong Spearman correlation (r = 0.829, P < .001). Phase 2 started after inversion of the sensors in the burr holes; there was a positive and moderately weak Spearman correlation (r = 0.262, P < .001). For phase 3, the sensors were returned to the first burr holes; the prototype behaved similarly to the reference sensor, presenting a positive and moderately strong Spearman correlation (r = 0.669, P < .001). CONCLUSION: A Wi-Fi ICP monitoring system was demonstrated in a comprehensive and feasible way. It was possible to observe, using smartphone, an adequate correlation regarding ICP variations. Further adaptations are already being developed.
引用
- da Silva Junior EB, Hamasaki EE, Smaili HY, et al. Fiber-Optic Intracranial Pressure Monitoring System Using Wi-Fi-An In Vivo Study. Neurosurgery. 2022. PMID: 36512829. doi:10.1227/neu.0000000000002250
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。