文献速递功能神经外科

Recharge PSYCH:精神疾病脑深部电刺激中可充电植入式脉冲发生器的研究

Recharge PSYCH: A Study on Rechargeable Implantable Pulse Generators in Deep Brain Stimulation for Psychiatric Disorders.
PubMed 文献 · World Neurosurg · 2022-11-09

Recharge PSYCH 研究向 21 例植入可充电植入式脉冲发生器的精神疾病脑深部电刺激患者发放标准化问卷,主要终点为充电便利性,中位评分 8.0 分(满分 10 分,评为「容易」),平均充电负担为每周 286 ± 22.4 分钟,33.3% 患者曾遇到设备无法成功充电,38.1% 患者治疗被意外中断。

要点

  • 精神疾病脑深部电刺激患者对可充电脉冲发生器充电过程的中位便利性评分为 8.0 分(满分 10 分,评为「容易」)。
  • 33.3% 患者曾遇到设备无法成功充电,38.1% 患者治疗被意外中断。
  • 平均充电负担为每周 286 ± 22.4 分钟。

核心结论

  • 精神疾病患者将充电过程评为「容易」,但与已发表的运动障碍脑深部电刺激患者数据相比,充电负担和使用相关并发症率显著更高。

研究设计与样本

  • Recharge PSYCH 研究是首个纳入接受不同可充电植入式脉冲发生器型号治疗的精神疾病脑深部电刺激患者的研究。
  • 研究向所有在研究时植入可充电植入式脉冲发生器的精神疾病脑深部电刺激患者发放标准化问卷。
  • 共有 21 例患者的数据集纳入分析。
  • 主要终点为充电便利性;次要终点为用户信心率、使用相关并发症率以及充电负担(定义为每周所需充电分钟数)。

主要发现

  • 调查时患者植入可充电植入式脉冲发生器的平均时间为 31.8 ± 22.4 个月。
  • 在植入可充电植入式脉冲发生器之前,患者接受过中位 3 次植入式脉冲发生器更换。
  • 充电过程的总体便利性中位评分为 8.0 分(满分 10.0 分),评为「容易」。
  • 33.3% 的患者经历过设备无法成功充电的情况;38.1% 的患者治疗被意外中断。
  • 平均充电负担为每周 286 ± 22.4 分钟。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Rechargeable implantable pulse generators (r-IPGs) for deep brain stimulation (DBS) promise longer battery life and fewer replacement surgeries versus non-rechargeable systems. Long-term data on the effects of recharging in patients who received DBS for psychiatric indications is limited. The Recharge PSYCH trial is the first study that included DBS patients with psychiatric disorders treated with different r-IPG models. METHODS: Standardized questionnaires were sent to all psychiatric DBS patients with an r-IPG implanted at the time of the study. The primary endpoint was convenience of recharging. Secondary endpoints were rate of user confidence and rate of usage-related complications, as well as charge burden (defined as minutes per week needed to recharge). RESULTS: Data sets of n = 21 patients were eligible for data analysis. At the time of the survey patients were implanted with the r-IPG for a mean 31.8 ± 22.4 months. Prior to being implanted with an r-IPG, patients had undergone a median of 3 IPG replacements. The overall convenience of the charging process was rated as "easy" with a median of 8.0 out of 10.0 points. 33.3% of patients experienced situations in which the device could not be successfully recharged. In 38.1% of patients, therapy with the r-IPG was interrupted unintentionally. The average charge burden was 286 ± 22.4 minutes per week. CONCLUSIONS: Patients with psychiatric disorders rated the recharging process as "easy", but with a significantly higher charge burden and usage-related complication rates compared to published data on movement disorder DBS patients.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Jakobs M, Hajiabadi MM, Aguirre-Padilla DH, et al. Recharge PSYCH: A Study on Rechargeable Implantable Pulse Generators in Deep Brain Stimulation for Psychiatric Disorders. World Neurosurg. 2022. PMID: 36368453. doi:10.1016/j.wneu.2022.11.017

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。