脑室腹腔分流管组装过程中手术器械与导管损伤的关系
本研究对85根利福平涂层导管和85根钡浸渍导管使用5种手术器械和2种技术进行失效测试,发现技术A(推入法)导管失败率显著低于技术B(拉套法)(5% vs 81%,p<0.001),带护套蚊式钳失败率最低,DeBakey钳最易导致失败。
要点
- 技术A(推入法)导管失败率5%显著低于技术B(拉套法)81%(p<0.001)。
- 带护套蚊式钳组装时导管失败率显著低于其他器械(p<0.0005)。
- DeBakey钳平均1.38次重复即导致失败,低于Gerald钳2.73次和bayonet钳3.25次。
核心结论
- 使用带护套蚊式钳和技术A组装分流管时导管失败率显著更低。
- DeBakey钳导致失败所需组装/拆卸重复次数最少,提示其损伤性最大。
研究设计与样本
- 对85根利福平涂层导管和85根钡浸渍导管使用5种手术器械和2种技术进行失效测试。
- 160根导管进行10次重复最大测试,5根钡浸渍和5根利福平涂层导管使用技术A进行100次重复最大测试。
主要发现
- 不同器械组装时导管失败率显著不同(p≤0.001)。
- 带护套蚊式钳导管失败率显著低于其他器械(p<0.0005)。
- 技术A导管失败率5%显著低于技术B的81%(p<0.001)。
- 钡浸渍导管与利福平涂层导管失败率无统计学差异(42% vs 44%,p=0.9),但100次重复最大测试中钡导管优于利福平导管(p=0.02)。
- 器械类型(p=0.0232)和导管类型(p=0.0096)均为导致导管失败所需组装/拆卸重复次数的显著因素,平均2.79次重复导致失败。
临床怎么用
- 作者认为使用带护套蚊式钳和技术A组装可降低导管失败率。
- 作者提示DeBakey钳损伤性最大。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Current surgical techniques use common surgical instruments for sterile shunt assembly. This study investigated the impact of using these techniques and surgical instruments on the mechanical integrity of the ventriculoperitoneal shunt system, specifically shunt catheters. METHODS: The authors conducted failure testing on 85 rifampin-coated catheters and 85 barium-impregnated catheters using 5 different surgical instruments and 2 different surgical techniques. In technique A, the distal end of the catheter was pushed onto the shunt valve inlet connector with the surgical instrument. In technique B, the catheter was pulled over the inlet connector. One hundred sixty catheters underwent 10-repetition-maximum testing, in which the catheter either failed before completion of 10 consecutive assembly/disassembly repetitions or the catheter completed 10 consecutive repetitions. The authors also conducted 100-repetition-maximum tests on 5 barium-impregnated and 5 rifampin-coated catheters using technique A. RESULTS: Catheter failure rates were significantly different among the different instruments used in assembly (p ≤ 0.001). Post hoc analysis showed that using mosquito forceps with shods resulted in a significantly lower catheter failure rate than the other instruments (p < 0.0005). The catheter failure rate of technique A was significantly lower than that of technique B (5% vs 81%, p < 0.001). There was no statistical difference between the failure rates of the barium and rifampin catheters (42% vs 44%, p = 0.9), but the barium catheters outperformed the rifampin catheters in the 100-repetition-maximum trials (p = 0.02). Instrument type (p = 0.0232) and catheter type (p = 0.0096) were both significant factors in determining the number of assembly/disassembly repetitions needed to cause catheter failure. It took an average of 2.79 repetitions of assembly/disassembly to cause catheter failure. DeBakey forceps had significantly lower mean repetitions to failure (mean 1.38) than the Gerald forceps without teeth (mean 2.73, p = 0.05) and bayonet (mean 3.25, p = 0.02). CONCLUSIONS: This study was the first of its kind to analyze how physical manipulation affects the mechanical integrity of ventriculoperitoneal shunt components. The authors demonstrated a significantly lower shunt catheter failure rate when mosquito forceps with shods and technique A were used in assembly. Moreover, the authors established that use of DeBakey forceps resulted in failure with fewer assembly/disassembly repetitions than use of the Gerald and bayonet forceps, suggesting that DeBakey forceps are the most damaging instrument.
引用
- Fisher WAM, Catalino MP, Woolard EA, et al. Surgical instruments and catheter damage during ventriculoperitoneal shunt assembly. J Neurosurg. 2022. PMID: 36242572. doi:10.3171/2022.8.JNS22746
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。