脑室外引流与腰大池引流的机械性并发症
单中心前瞻性队列研究纳入2020年1月至2021年3月需脑脊液分流的61例急性脑积水患者,21例(39%)发生机械性并发症,以堵管最常见。
要点
- 61例患者中21例(39%)发生机械性并发症,堵管最常见。
核心结论
- 引流装置影响脑室外引流相关并发症的发生,初步发现提示选择螺栓固定脑室外引流和电机辅助引流可减少引流相关机械性失败。
- 作者认为需开展比较引流装置的随机对照试验以证实这些发现。
研究设计与样本
- 单中心前瞻性队列研究,纳入2020年1月至2021年3月因急性脑积水需脑脊液分流的61例患者。
- 使用3种引流类型(腰大池引流、隧道式脑室外引流、螺栓固定脑室外引流)和2种收集系统(LiquoGuard 7和AccuDrain),在手术室或重症监护室置入。
- 采用单变量分析探索数据趋势,随后用重复测量混合效应回归确定引流装置对机械性失败的独立影响。
主要发现
- 61例患者中21例(39%)发生机械性并发症,堵管最常见。
临床怎么用
- 作者认为选择螺栓固定脑室外引流和电机辅助引流可减少引流相关机械性失败。
- 作者建议开展比较引流装置的随机对照试验以证实这些发现。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: External ventricular drain (EVD) and lumbar drain insertion are 2 of the most commonly performed neurosurgical procedures worldwide for acute hydrocephalus. Mechanical complications, such as obstruction or cerebrospinal fluid (CSF) leakage, are often seen and may contribute toward significant patient morbidity. Different CSF drainage methods are advocated to reduce the incidence of complications, but evidence regarding comparative effectiveness is limited. METHODS: In this single-center prospective cohort study, the incidence of mechanical complications and associated interventional factors, including choice of drain, collecting system, and location, were studied in patients requiring CSF diversion for acute hydrocephalus. Univariate analyses were performed to explore trends within the data, followed by a repeated-measures mixed-effects regression to determine the independent influence of drain device on mechanical failure. RESULTS: Sixty-one patients required CSF diversion between January 2020 and March 2021, via 3 different drain types (lumbar drain, tunneled, and bolted EVD) and 2 collection systems (LiquoGuard 7 and AccuDrain), performed in either theater or intensive care. Twenty-one patients (39%) experienced a mechanical complication, with blockage being the most common. Multivariate analyses showed that bolted EVDs (odds ratio, 0.08; confidence interval, 0.01-0.58) and LiquoGuards (OR, 0.23; CI, 0.08-0.69) were significantly associated with fewer mechanical complications compared with tunneled EVDs and gravity-based collection systems, respectively (P ≤ 0.01). DISCUSSION: Drain device has an influence on the occurrence of EVD-related complications. These preliminary findings suggest that choosing bolted EVDs and motor-assisted drainage can reduce drain-associated mechanical failure. A randomized controlled trial comparing drain devices is required to confirm these findings.
引用
- Pandit AS, Palasz J, Nachev P, et al. Mechanical Complications of External Ventricular and Lumbar Drains. World Neurosurg. 2022. PMID: 35787961. doi:10.1016/j.wneu.2022.06.127
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。