比较颅骨密度比与其他颅骨指标对磁共振引导聚焦超声丘脑毁损术治疗震颤的影响
该回顾性研究纳入2017至2021年在本中心接受MRgFUS丘脑毁损术的62例特发性震颤患者,评估颅骨密度比(SDR)及其他颅骨指标与治疗参数的相关性,并应用机器学习预测超声参数。SDR与最大功率(ρ=-0.626,p<0.001)和最大能量(ρ=-0.680,p<0.001)相关性最高,机器学习预测最大功率和能量的准确率分别约80%和55%,预测平均最高温度的准确率约95%。
要点
- SDR与最大功率(ρ=-0.626,p<0.001)和最大能量(ρ=-0.680,p<0.001)相关性最高。
- 机器学习预测最大功率准确率约80%、最大能量约55%、平均最高温度约95%。
- 作者认为SDR是单独使用时最佳的治疗参数指标之一,其他算法有待更多数据探索。
核心结论
- 作者比较多种颅骨指标与SDR,显示SDR是单独使用时最佳的治疗参数指标之一;并提出其他机器学习算法,待获得更多数据时可探索以提高准确性。
研究设计与样本
- 回顾性研究连续纳入2017至2021年在本中心接受MRgFUS丘脑毁损术的特发性震颤患者,共62例。
主要发现
- 平均年龄77.1岁(SD 9.2),78%的治疗(49/63)在男性中进行,平均SDR为0.51(SD 0.10)。
- SDR与最大功率相关性最高(ρ=-0.626,p<0.001),局部SDR值≤0.8组比例也有ρ=+0.626,p<0.001;SDR与最大能量相关性最高(ρ=-0.680,p<0.001)。
- 机器学习从局部和整体SDR预测最大功率和所需能量的能力中等(最大功率准确率约80%,最大能量约55%),预测平均最高温度的能力高(准确率约95%)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: One of the key metrics that is used to predict the likelihood of success of MR-guided focused ultrasound (MRgFUS) thalamotomy is the overall calvarial skull density ratio (SDR). However, this measure does not fully predict the sonication parameters that would be required or the technical success rates. The authors aimed to assess other skull characteristics that may also contribute to technical success. METHODS: The authors retrospectively studied consecutive patients with essential tremor who were treated by MRgFUS at their center between 2017 and 2021. They evaluated the correlation between the different treatment parameters, particularly maximum power and energy delivered, with a range of patients' skull metrics and demographics. Machine learning algorithms were applied to investigate whether sonication parameters could be predicted from skull density metrics alone and whether including combined local transducer SDRs with overall calvarial SDR would increase model accuracy. RESULTS: A total of 62 patients were included in the study. The mean age was 77.1 (SD 9.2) years, and 78% of treatments (49/63) were performed in males. The mean SDR was 0.51 (SD 0.10). Among the evaluated metrics, SDR had the highest correlation with the maximum power used in treatment (ρ = -0.626, p < 0.001; proportion of local SDR values ≤ 0.8 group also had ρ = +0.626, p < 0.001) and maximum energy delivered (ρ = -0.680, p < 0.001). Machine learning algorithms achieved a moderate ability to predict maximum power and energy required from the local and overall SDRs (accuracy of approximately 80% for maximum power and approximately 55% for maximum energy), and high ability to predict average maximum temperature reached from the local and overall SDRs (approximately 95% accuracy). CONCLUSIONS: The authors compared a number of skull metrics against SDR and showed that SDR was one of the best indicators of treatment parameters when used alone. In addition, a number of other machine learning algorithms are proposed that may be explored to improve its accuracy when additional data are obtained. Additional metrics related to eventual sonication parameters should also be identified and explored.
引用
- Yuen J, Goyal A, Kaufmann TJ, et al. Comparison of the impact of skull density ratio with alternative skull metrics on magnetic resonance-guided focused ultrasound thalamotomy for tremor. J Neurosurg. 2022. PMID: 35901729. doi:10.3171/2022.5.JNS22350
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。