新型机器视觉辅助立体定向神经外科用于脑室外引流置管的初步研究
该初步研究探讨了利用机器视觉技术进行患者注册和立体定向脑室外引流(EVD)的手术导航机器人NaoTrac的效能。14例病例均通过导管引流脑脊液,NaoTrac在13例中一次尝试将导管置入额角,12例成功引流。平均患者注册时间142.8秒,平均靶点偏差1.68毫米,平均角度偏差1.99度,均在最小组织损伤的可接受容差内,无出血或术中并发症。
要点
- NaoTrac在13例临床病例中完成无框架导管置入,展示其准确性和安全性。
- 14例均通过导管引流脑脊液,13例一次尝试置入额角,12例成功引流。
- 平均靶点偏差1.68毫米,平均角度偏差1.99度,无出血或术中并发症。
核心结论
- 机器视觉启发的患者注册可行且快速,NaoTrac在13例临床病例的无框架导管置入中展示了准确性和安全性。
- 其他立体定向神经外科手术如立体定向活检、深部电极置入、脑深部电刺激电极定位和神经内镜也可能受益于NaoTrac的辅助。
研究设计与样本
- 初步研究,共14例病例,使用手术导航机器人NaoTrac(Brain Navi Biotechnology Co., Ltd.)进行患者注册和立体定向脑室外引流。
- 每例均获取术前和术后计算机断层扫描(CT),术前CT用于规划靶点和轨迹,术后CT用于准确性测量。
主要发现
- 全部14例均通过导管引流脑脊液。
- NaoTrac在13例中一次尝试将导管置入额角,并在12例中成功引流脑脊液。
- 无任何病例出现出血或术中并发症。
- 平均患者注册时间为142.8秒。
- 平均靶点偏差为1.68毫米,平均角度偏差为1.99度,均在最小组织损伤的可接受容差内。
关键数值与推荐
- 原文未报告
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The usage of machine vision technologies for image-based analysis and inspection is increasing. With the advent of the ability to process high-dimension data instantly, the possibilities of machine vision multiply exponentially. Robots now use this technology to assist in surgery. OBJECTIVE: The aim of this study is to explore the efficacy of Surgical Navigation Robot NaoTrac (Brain Navi Biotechnology Co., Ltd.), which utilizes machine vision-inspired technology for patient registration and stereotactic external ventricular drainage (EVD) by the robotic arm. METHODS: Preoperative and postoperative computed tomography (CT) scans were acquired for each case. The surgeons planned the targets and trajectories with the preoperative CT images. The postoperative CT images were utilized in the accuracy measurements. RESULTS: All 14 cases had cerebrospinal fluid drained through the catheter. The NaoTrac placed the catheter into the frontal horn in one attempt in 13 cases and was able to drain CSF in 12 cases. Not a single case had any bleeding or intraoperative complications. The average time spent on the patient registration was 142.8 s. The mean target deviation was 1.68 mm, and the mean angular deviation was 1.99°, all within the accepted tolerance for minimal tissue damage. CONCLUSION: The results of this report demonstrate that machine vision-inspired patient registration is feasible and fast. NaoTrac has demonstrated its accuracy and safety in performing frameless catheter placement in 13 clinical cases. Other stereotactic neurosurgical operations such as stereotactic biopsy, depth electrode placement, deep brain stimulation electrode positioning, and neuroendoscopy may also be benefited from the assistance of NaoTrac.
引用
- Chiu TL, Lin SZ, Ahmed T, et al. Pilot study of a new type of machine vision-assisted stereotactic neurosurgery for EVD placement. Acta Neurochir (Wien). 2022. PMID: 35788905. doi:10.1007/s00701-022-05287-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。