立体脑电图与脑深部电刺激中无框架瞬态基准点配准的准确性与效率
该研究回顾性分析了2015年3月至2020年3月同一机构88例(621个电极)的脑深部电刺激(DBS)或立体脑电图(SEEG)手术,比较无框架瞬态基准点配准(FTFR,Neurolocate系统)与框架基准点配准(FBR)的准确性和效率。结果显示FTFR配准精度更高(中位0.1 mm vs 1.3 mm,p=0.04)且配准时间更短(中位84分钟 vs 110.5分钟,p=0.02),径向入点和欧几里得靶点植入误差无显著差异。作者认为FTFR可能缩短手术时间并保持高精度,但需未来研究进一步验证。
要点
- FTFR配准精度中位0.1 mm,显著优于FBR的1.3 mm(p=0.04)。
- FTFR配准时间中位84分钟,显著短于FBR的110.5分钟(p=0.02)。
- 两种方法在径向入点和欧几里得靶点植入误差上无显著差异。
核心结论
- FTFR with the Neurolocate system may decrease operative time while maintaining high accuracy, despite an initial surgeon learning curve.
- It should be investigated in future studies to continue to improve stereotactic accuracies in neurosurgery.
研究设计与样本
- 回顾性图表和影像审查,纳入作者所在机构2015年3月至2020年3月5年期间连续88例植入DBS或SEEG的手术(涉及621个电极)。
- SEEG手术包括38例患者和525个植入电极(294个使用FBR,231个使用FTFR)。
- DBS手术包括50例患者和96个植入电极(65个使用FBR,31个使用FTFR)。
主要发现
- 总体FTFR配准精度(中位0.1 mm,IQR 0-0.4 mm)显著优于FBR(中位1.3 mm,IQR 0.9-1.5 mm;p=0.04)。
- FTFR配准持续时间(中位84分钟,IQR 77.3-95.3分钟)显著短于FBR(中位110.5分钟,IQR 107.3-138分钟;p=0.02)。
- 检查每种方法的径向入点和欧几里得靶点植入误差时未发现显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Stereotactic surgical methods continue to advance technologically. Frameless transient fiducial registration (FTFR) systems have been developed and avoid the need to move or position a patient in a frame after already receiving registration imaging. One such system, Neurolocate, has recently become available as a robotic attachment for the Neuromate stereotactic robot. This study is the largest in the literature to evaluate the accuracy of frameless registration using Neurolocate versus frame-based registration (FBR) methods in both deep brain stimulation (DBS) and stereoelectroencephalography (SEEG). Additionally, the authors sought to reevaluate factors affecting accuracy in both procedures. METHODS: This study was a retrospective chart and imaging review of 88 consecutive procedures (involving 621 electrodes) implanting either DBS or SEEG at the authors' institution over a 5-year period from March 2015 to March 2020. Registration duration, radial target entry point, and Euclidean target implantation accuracies, as well as factors affecting accuracy, were recorded for each patient. RESULTS: SEEG procedures included 38 patients and 525 implanted electrodes (294 using FBR and 231 using FTFR). DBS procedures included 50 patients and 96 implanted electrodes (65 using FBR and 31 using FTFR). Overall, FTFR registration was significantly more accurate (median 0.1 mm, IQR 0-0.4 mm) compared with FBR (median 1.3 mm, IQR 0.9-1.5 mm; p = 0.04). Likewise, FTFR had a significantly shorter duration of registration (median 84 minutes, IQR 77.3-95.3 minutes) when compared with FBR (median 110.5 minutes, IQR 107.3-138 minutes; p = 0.02). No significant differences were found when examining the radial entry point and Euclidean target implantation errors of each method. CONCLUSIONS: FTFR with the Neurolocate system represents a technique that may decrease operative time while maintaining the high accuracy previously demonstrated by other stereotactic methods, despite an initial surgeon learning curve. It should be investigated in future studies to continue to improve stereotactic accuracies in neurosurgery.