血液透析患者革兰阴性菌血流感染的多中心暴发
共58例病例,48例需住院,以粘质沙雷菌和铜绿假单胞菌为主。使用中心静脉导管透析的比值比显著升高(匹配OR 54.32,95%CI下限12.19)。暴发与墙盒污染相关,为先前未描述的感染源。
要点
- 中心静脉导管使用与感染风险显著相关(匹配OR 54.32)。
- 墙盒可滋生粘质沙雷菌和铜绿假单胞菌生物膜。
- 需关注透析单元墙盒的清洁消毒与手卫生。
核心结论
- 本次暴发与墙盒相关,墙盒是患者照护环境中先前未描述的污染液体和生物膜来源。
主要发现
- 共发生58例革兰阴性菌血流感染,48例(83%)需住院。
- 主要病原体为粘质沙雷菌(n=21)和铜绿假单胞菌(n=12)。
- 病例使用中心静脉导管透析的比值比高于对照(匹配OR 54.32,95%CI下限12.19)。
- 墙盒环境样本检出粘质沙雷菌和铜绿假单胞菌,且与病例分离株高度相关。
临床怎么用
- 应关注透析单元墙盒的污染与生物膜,加强清洁消毒和手卫生。
- 减少不必要的中心静脉导管使用以降低感染风险。
局限性
- 可供检测的患者分离株有限。
- 现场调查在感染高峰之后进行。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
RATIONALE & OBJECTIVE: Contaminated water and other fluids are increasingly recognized to be associated with health care-associated infections. We investigated an outbreak of Gram-negative bloodstream infections at 3 outpatient hemodialysis facilities. STUDY DESIGN: Matched case-control investigations. SETTING & PARTICIPANTS: Patients who received hemodialysis at Facility A, B, or C from July 2015 to November 2016. EXPOSURES: Infection control practices, sources of water, dialyzer reuse, injection medication handling, dialysis circuit priming, water and dialysate test findings, environmental reservoirs such as wall boxes, vascular access care practices, pulsed-field gel electrophoresis, and whole-genome sequencing of bacterial isolates. OUTCOMES: Cases were defined by a positive blood culture for any Gram-negative bacteria drawn July 1, 2015 to November 30, 2016 from a patient who had received hemodialysis at Facility A, B, or C. ANALYTICAL APPROACH: Exposures in cases and controls were compared using matched univariate conditional logistic regression. RESULTS: 58 cases of Gram-negative bloodstream infection occurred; 48 (83%) required hospitalization. The predominant organisms were Serratia marcescens (n=21) and Pseudomonas aeruginosa (n=12). Compared with controls, cases had higher odds of using a central venous catheter for dialysis (matched odds ratio, 54.32; lower bound of the 95% CI, 12.19). Facility staff reported pooling and regurgitation of waste fluid at recessed wall boxes that house connections for dialysate components and the effluent drain within dialysis treatment stations. Environmental samples yielded S marcescens and P aeruginosa from wall boxes. S marcescens isolated from wall boxes and case-patients from the same facilities were closely related by pulsed-field gel electrophoresis and whole-genome sequencing. We identified opportunities for health care workers' hands to contaminate central venous catheters with contaminated fluid from the wall boxes. LIMITATIONS: Limited patient isolates for testing, on-site investigation occurred after peak of infections. CONCLUSIONS: This large outbreak was linked to wall boxes, a previously undescribed source of contaminated fluid and biofilms in the immediate patient care environment.
引用
- Novosad SA, Lake J, Nguyen D, et al. Multicenter Outbreak of Gram-Negative Bloodstream Infections in Hemodialysis Patients. Am J Kidney Dis. 2019. PMID: 31375298. doi:10.1053/j.ajkd.2019.05.012
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。