生物电阻抗与透析治疗儿童及青少年的液体状态

Bioimpedance and Fluid Status in Children and Adolescents Treated With Dialysis.
PubMed 文献 · Am J Kidney Dis · 2017-01-13

本研究为诊断试验准确性研究,纳入16例接受透析的儿童及青少年,比较生物电阻抗与氘稀释法测总水量、溴稀释法测细胞外液量。结果显示生物电阻抗无系统偏倚但精度不足,不能精确估计总水量和细胞外液量。

要点

  • 生物电阻抗测总水量与氘稀释法平均偏差为-0.09 L,95%一致性界限-2.1至1.9 L。
  • 生物电阻抗测细胞外液量与溴稀释法平均偏差为+0.6 L,95%一致性界限-2.3至3.5 L。
  • 作者认为生物电阻抗不能精确估计透析儿童的总水量和细胞外液量。

核心结论

  • 与稀释技术相比,生物电阻抗测量无偏倚但不精确,不能精确估计接受透析儿童的总水量和细胞外液量。

研究设计与样本

  • 诊断试验准确性研究,纳入16例来自意大利米兰和荷兰阿姆斯特丹两家医院的年轻透析患者(血液透析前或腹膜透析治疗后)。

主要发现

  • 生物电阻抗与氘稀释法测平均总水量分别为19.2±8.7 L和19.3±8.3 L,平均偏差-0.09 L(95%一致性界限-2.1至1.9 L)。
  • 生物电阻抗与溴稀释法测平均细胞外液量分别为8.9±4.0 L和8.3±3.3 L,平均偏差+0.6 L(95%一致性界限-2.3至3.5 L)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 参与者在无研究者直接监督下在家中饮用氘水。
  • 患者数量少。
  • 未对个体患者进行重复测量。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Assessment of hydration status in patients with chronic kidney failure treated by dialysis is crucial for clinical management decisions. Dilution techniques are considered the gold standard for measurement of body fluid volumes, but they are unfit for day-to-day care. Multifrequency bioimpedance has been shown to be of help in clinical practice in adults and its use in children and adolescents has been advocated. We investigated whether application of multifrequency bioimpedance is appropriate for total-body water (TBW) and extracellular water (ECW) measurement in children and adolescents on dialysis therapy. STUDY DESIGN: A study of diagnostic test accuracy. SETTING & PARTICIPANTS: 16 young dialysis patients (before a hemodialysis session or after peritoneal dialysis treatment) from the Ca' Granda Ospedale Maggiore Policlinico, Milan, Italy, and the Emma Children's Hospital-Academic Medical Center, Amsterdam, the Netherlands. INDEX TEST: TBW and ECW volumes assessed by multifrequency bioimpedance. REFERENCE TESTS: TBW and ECW volumes measured by deuterium and bromide dilution, respectively. RESULTS: Mean TBW volumes determined by multifrequency bioimpedance and deuterium dilution were 19.2±8.7 (SD) and 19.3±8.3L, respectively; Bland-Altman analysis showed a mean bias between the 2 methods of -0.09 (95% limits of agreement, -2.1 to 1.9) L. Mean ECW volumes were 8.9±4.0 and 8.3±3.3L measured by multifrequency bioimpedance and bromide dilution, respectively; mean bias between the 2 ECW measurements was +0.6 (95% limits of agreement, -2.3 to 3.5). LIMITATIONS: Participants ingested the deuterated water at home without direct supervision by investigators, small number of patients, repeated measurements in individual patients were not performed. CONCLUSIONS: Multifrequency bioimpedance measurements were unbiased but imprecise in comparison to dilution techniques. We conclude that multifrequency bioimpedance measurements cannot precisely estimate TBW and ECW in children receiving dialysis.

来源:Am J Kidney Dis,查看原文。版权归原作者所有。

引用

  1. Milani GP, Groothoff JW, Vianello FA, et al. Bioimpedance and Fluid Status in Children and Adolescents Treated With Dialysis. Am J Kidney Dis. 2017. PMID: 28089477. doi:10.1053/j.ajkd.2016.10.023