一次性试纸用于维持性血液透析患者血钾检测的临床评估

Clinical Evaluation of a Disposable Test Strip for Potassium Level Measurement in Patients Treated with Maintenance Hemodialysis.
PubMed 文献 · Clin J Am Soc Nephrol · 2025-12-05

本研究评估了一次性钾试纸在实验条件和维持性血液透析患者中的性能。试纸在2-9 mM钾浓度范围内变异系数<5%,毛细血管血钾与中心实验室血清钾的Passing-Bablok回归斜率为1.04(95%CI 0.97-1.12),显示高度一致,可用于临床自我监测。

要点

  • 一次性钾试纸在2-9 mM范围内变异系数<5%,与标准溶液高度一致。
  • 40例维持性血液透析患者中,毛细血管血钾与中心实验室血清钾高度一致。
  • 试纸法可用于血钾自我监测,但需注意研究样本量较小。

核心结论

  • 一次性钾试纸POCT可用于维持性血液透析患者血钾的自我监测。

研究设计与样本

  • 实验研究:测试不同钾浓度溶液(含或不含干扰电解质)中一次性钾试纸的性能。
  • 临床研究:在40例维持性血液透析患者中,比较试纸法与中心实验室设备测定血钾的一致性,使用Passing-Bablok回归和Bland-Altman分析。

主要发现

  • 试纸法测钾与电感耦合等离子体原子发射光谱法高度一致。
  • 在2至9 mM钾浓度范围内,试纸的变异系数<5%。
  • 毛细血管血试纸法测钾与中心实验室血清钾高度一致,Passing-Bablok回归斜率为1.04(95%CI 0.97-1.12)。
  • 其他类型血样的分析也显示类似结果。

临床怎么用

  • 一次性钾试纸可用于维持性血液透析患者血钾的自我监测。

局限性

  • 临床研究样本量仅为40例维持性血液透析患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

KEY POINTS: Potassium level test strip showed excellent performance with accuracy and reliability in experimental setting with standard electrolyte solutions. Potassium levels obtained with test strips yielded satisfactory agreement with those measured by central laboratory equipment. BACKGROUND: Hyperkalemia is a common electrolyte imbalance leading to an increased risk of serious cardiac dysrhythmias and mortality in patients undergoing hemodialysis. A convenient point-of-care testing (POCT) of potassium levels is considered to improve patient outcomes. However, POCT for potassium level using capillary blood with satisfactory performance is not available. METHODS: We tested the performance of disposable test strips for potassium level using various concentrations of potassium solutions with or without interfering electrolytes in an experimental setting. Subsequently, we examined the agreement between potassium levels measured by disposable test strips and those measured by central laboratory equipment using various types of blood samples, including capillary blood obtained by finger prick, in 40 patients undergoing maintenance hemodialysis. In the analysis, Passing-Bablok regression and Bland-Altman analysis were used. RESULTS: Potassium concentrations measured by disposable test strips showed a high degree of agreement with potassium levels achieved using inductively coupled plasma atomic emission spectroscopy. The coefficient of variations for disposable test strips were <5% across potassium concentrations ranging from 2 to 9 mM. Furthermore, potassium levels measured by disposable strips from capillary blood demonstrated a high degree of agreement with potassium levels obtained by central laboratory equipment using serum, showing the slope achieved from Passing-Bablok regression was 1.04 (95% confidence interval, 0.97 to 1.12). Additional analysis with various blood samples also showed similar results. CONCLUSIONS: POCT for potassium level using the disposable test strip could be used for self-monitoring of blood potassium levels in clinical practice.

来源:Clin J Am Soc Nephrol,查看原文。版权归原作者所有。