新型比色与光散射法在床旁识别和管理腹膜透析相关腹膜炎
本研究开发了两种床旁检测方法:基于四唑鎓的化学报告系统用于检测细菌污染及万古霉素敏感性,新型前向光散射设备QuickCheck用于即时定量细胞计数。采用患者透出液样本比较,QuickCheck准确度高(94%),精密度最佳(变异系数<4%),性能与流式细胞术相似。两种互补方法可在床旁辅助腹膜炎诊断。
要点
- 四唑鎓法10小时提供非定量细菌检测及万古霉素敏感性结果。
- QuickCheck光散射法即时准确计数透出液白细胞,准确度94%。
- 两种床旁方法互补,可辅助腹膜透析相关腹膜炎的早期诊断。
核心结论
- 两种互补的新型床旁方法可提供辅助腹膜炎诊断的信息:四唑鎓法10小时孵育后提供抗生素敏感性,光散射法即时提供准确的白细胞总数。
研究设计与样本
- 开发两种床旁检测方法:基于四唑鎓的化学报告系统检测细菌污染及万古霉素敏感性;新型前向光散射设备QuickCheck即时定量细胞计数。
- 使用患者透出液样本比较光散射白细胞计数与分光光度法、血细胞计数板法和流式细胞术(Sysmex)的技术性能。
主要发现
- 四唑鎓法可检测并正确区分实验室分离菌株,10小时提供非定量结果。
- QuickCheck准确度高(94%),精密度最佳(变异系数<4%),偏差极小,总体性能与流式细胞术相似。
关键数值与推荐
- QuickCheck准确度为94%,变异系数<4%。
临床怎么用
- 四唑鎓法可在床旁10小时孵育后提供细菌检测及万古霉素敏感性信息。
- QuickCheck可在床旁即时提供准确的透出液白细胞总数,辅助腹膜炎诊断。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Peritoneal dialysis (PD)-related peritonitis (PDRP) is a common cause of transfer to hemodialysis, patient morbidity, and is a risk factor for mortality. Associated patient anxiety can deter selection of PD for renal replacement therapy. Diagnosis relies on hospital laboratory tests; however, this might be achieved earlier if such information was available at the point-of-care (POC), thereby significantly improving outcomes. The presence of culturable microbes and the concentration of leukocytes in effluent both aid peritonitis diagnosis, as specified in the International Society for Peritoneal Dialysis (ISPD) diagnostic guidelines. Here, we report the development of 2 new methods providing such information in simple POC tests. METHODS: One approach uses a tetrazolium-based chemical reporting system, primarily focused on detecting bacterial contamination and associated vancomycin-sensitivity. The second approach uses a novel forward light-scatter device (QuickCheck) to provide an instant quantitative cell count directly from PD patient effluent. RESULTS: The tetrazolium approach detected and correctly distinguished laboratory isolates, taking 10 hours to provide non-quantitative results. We compared the technical performance of the light scatter leukocyte counting approach with spectrophotometry, hemocytometer counting and flow cytometry (Sysmex) using patient effluent samples. QuickCheck had high accuracy (94%) and was the most precise (coefficient of variation <4%), showing minimal bias, overall performing similarly to flow cytometry. CONCLUSION: These complementary new approaches provide a simple means to obtain information to assist diagnosis at the POC. The first provides antibiotic sensitivity following 10 hours incubation, whereas the second optical approach (QuickCheck), provides instant accurate total leukocyte count.
引用
- Govindji-Bhatt N, Kennedy SM, Barker MG, et al. Novel Colorimetric and Light Scatter Methods to Identify and Manage Peritoneal Dialysis-Associated Peritonitis at the Point-of-Care. Kidney Int Rep. 2023. PMID: 38481507. doi:10.1016/j.ekir.2023.12.021
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。