微芯片与压电晶体床旁超声在急性肾损伤患者肺及下腔静脉评估中的比较
结果显示两种技术对肺B线(ρ=0.96和0.93)和IVCi(ρ=0.70和0.87)均有强正相关,Bland-Altman分析显示一致性良好。微芯片超声便携、易用且成本效益高,可提供与压电晶体超声相当的容量评估参数。
要点
- 微芯片与压电晶体超声评估肺B线在T0和T60均呈强相关(ρ=0.96和0.93)。
- 两种技术评估IVCi在T0和T60的相关性分别为ρ=0.70和ρ=0.87。
- Bland-Altman分析显示两种方法在IVCi和B线量化上一致性良好。
核心结论
- 微芯片与压电晶体床旁超声在评估急性肾损伤患者的肺B线和下腔静脉动态方面具有良好的相关性和一致性。
- 微芯片超声便携、易用且成本效益高,可提供与压电晶体超声相当的床旁容量评估参数。
研究设计与样本
- 采用随机顺序在肾脏替代治疗前(T0)和治疗开始后60分钟(T60)分别使用两种超声技术采集参数。
- 肺超声量化8个象限的B线,下腔静脉塌陷指数(IVCi)通过呼吸周期内最大和最小直径计算。
主要发现
- 肺B线在T0和T60的相关系数分别为ρ=0.96和ρ=0.93(均P<0.001)。
- IVCi在T0和T60的相关系数分别为ρ=0.70和ρ=0.87(均P<0.001)。
- IVCi在T0的Bland-Altman偏倚为+2.69(SD:10.6),95%CI: -18.13至+23.52;T60偏倚为3.28(SD:10.23),95%CI: -16.77至+23.34。
- B线量化在T0的偏倚为-0.3(SD:2.73),95%CI: -5.66至+5.06;T60偏倚为0.2(SD:3.23),95%CI: -6.14至+6.54。
临床怎么用
- 微芯片床旁超声可用于急性肾损伤危重患者的肺B线和下腔静脉动态评估,结果与压电晶体超声相当。
- 其便携性和易用性可能有助于在资源有限或床旁快速评估场景中推广使用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Point-of-care ultrasonography (POCUS) has emerged as an important tool for examining critically ill patients. POCUS devices have become progressively smaller and more accessible, transforming medical practice, and reducing costs. One technological breakthrough was the development of ultraportable scanners with microchip technology, which utilize a probe connected to a smartphone or tablet and incorporate a mobile application that employs artificial intelligence to assist in the interpretation of acquired images. The aim of this study was to compare the accuracy of ultrasound microchip technology with traditional piezoelectric crystal ultrasonography. We analyzed 2 volume status parameters in intensive care unit (ICU) patients with acute kidney injury (AKI) who were receiving kidney replacement therapy (KRT). These parameters were the extravascular pulmonary water, using the lung B-lines; and the inferior vena cava collapsibility index (IVCi). METHODS: Fifty critically ill patients met the study criteria. Lung POCUS quantified B-lines in 8 quadrants. The IVCi was measured using the maximum and minimum diameters during a respiratory cycle. Both technologies were sequentially employed in a randomized fashion to acquire the parameters at 2 different time points: before the initiation of KRT (T0) and 60 minutes after the procedure had commenced (T60). We calculated the correlation and agreement between the 2 ultrasound scanner modalities. RESULTS: The correlation between the 2 technologies for evaluating lung B-lines showed strong positive coefficients, ⍴ = 0.96 and ⍴ = 0.93 at T0 and T60, respectively (P < 0.001 for both). The correlation for IVCi was ρ = 0.70 and ρ = 0.87 at T0 and T60, respectively (P < 0.001 for both). The Bland-Altman plots showed agreement between ultrasound methods for IVCi calculation and B-line quantification. For IVCi calculation at T0, bias was +2.69 (SD: 10.6) (95% confidence interval [CI]: -18.13 to +23.52); at T60, bias was 3.28 (SD: 10.23) (95% CI: -16.77 to +23.34). For B-line quantification, the analysis yielded a bias of -0.3 (SD: 2.73), with a 95% CI of -5.66 to +5.06 at T0; and a bias of 0.2 (SD: 3.23), with a 95% CI of -6.14 to +6.54 at T60. CONCLUSION: Our study observed a good correlation and agreement between microchip and piezoelectric-based ultrasound modalities in evaluating the presence of pulmonary B-lines and inferior vena cava (IVC) dynamics in patients with AKI. Microchip ultrasound, in addition to being portable, user-friendly, and cost-effective, consistently delivers bedside parameters for volume assessment that are comparable to those obtained through POCUS using piezoelectric transducer-based ultrasonography.
引用
- Soares DM, de Souza Mendes R, Rocco Suassuna JH, et al. Microchip Versus Piezoelectric Point of Care Ultrasonography for Pulmonary and Vena Cava Evaluation in Patients With Acute Kidney Injury. Kidney Int Rep. 2023. PMID: 38344740. doi:10.1016/j.ekir.2023.11.019