人工智能辅助数字形态学系统用于骨髓涂片检查的性能评估
本研究对158例患者骨髓涂片样本,比较CellaVision BMA应用(自动预分类后专家审核)与光学显微镜的细胞分类计数,采用Pearson相关、Passing-Bablok回归和Bland-Altman分析。结果显示多数细胞类型相关性良好(r=0.85-0.99),但嗜碱性粒细胞和浆细胞一致性较弱,20%原始细胞阈值下灵敏度为77.3%,专家审核仍必要。
要点
- CellaVision BMA应用在多数有核细胞分类计数上与光学显微镜可比。
- 嗜碱性粒细胞和浆细胞一致性较弱,需专家审核。
- 20%原始细胞阈值下灵敏度77.3%,临床决策需谨慎。
核心结论
- 按预期工作流程使用CellaVision BMA应用时,其分类计数在多数有核细胞上与常规显微镜可比,但嗜碱性粒细胞和浆细胞一致性较弱,专家审核与验证仍必要。
研究设计与样本
- 单中心研究,分析158例患者骨髓涂片样本,比较光学显微镜与CellaVision BMA应用(自动预分类后专家审核)的分类计数。
主要发现
- 除嗜碱性粒细胞(r=0.68)外,所有细胞群Pearson相关系数为0.85至0.99。
- Passing-Bablok回归显示红系、粒系、嗜酸性粒细胞和单核细胞无显著系统或比例偏倚;原始细胞有小的系统差异,淋巴细胞有小的比例差异,浆细胞和嗜碱性粒细胞显著偏离线性。
- Bland-Altman平均差异在所有细胞群中范围为−1.4%至1.2%。
- 在原始细胞和浆细胞的临床相关决策阈值处,两种方法总体一致性为96.8%至99.4%,但20%原始细胞阈值下灵敏度为77.3%。
临床怎么用
- 该系统的分类计数可作为常规显微镜的辅助,但嗜碱性粒细胞和浆细胞结果需专家复核,原始细胞阈值判断应谨慎。
局限性
- 单中心系列研究,结果需进一步评估该及类似系统。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
ABSTRACT Introduction Morphological examination of bone marrow aspirates (BMA) remains essential for the diagnosis of hematologic diseases, but conventional microscopy is labor-intensive and subject to interobserver variability. Methods A total of 158 samples from patients were analyzed. Differential counts obtained by optical microscopy were compared with those of the CellaVision BMA Application used according to its intended workflow of automated preclassification followed by expert review. The two methods were compared by Pearson correlation, Passing–Bablok regression, and Bland–Altman analysis, both overall and within the blast and plasma cells intervals that bracket the diagnostic thresholds. Results Pearson correlation coefficients ranged from 0.85 to 0.99 for all populations except basophils (r = 0.68). Passing–Bablok regression showed no significant systematic or proportional bias for erythroblasts, granulocytic cells, eosinophils, and monocytes; a small systematic difference was observed for blasts, a small proportional difference for lymphocytes, and significant deviations from linearity for plasma cells and basophils. Bland–Altman mean differences ranged from −1.4% to 1.2% across all populations. At the clinically relevant decision thresholds for blasts and plasma cells, overall agreement between the two methods ranged from 96.8% to 99.4%, although sensitivity at the 20% blast threshold was 77.3%. Conclusion Used according to its intended workflow, the CellaVision BMA Application produced differential counts comparable with conventional microscopy for most nucleated cells, but agreement was weaker for basophils and plasma cells. Expert review and validation remain necessary. These findings, obtained in a single-center series, provide a basis for further evaluation of this and similar systems.
引用
- International Journal of Laboratory Hematology:Performance Evaluation of an Artificial Intelligence‐Assisted Digital Morphology System for Bone Marrow Aspirate Examination
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。