文献速递临床微生物质量管理与危急值

MicroScan WalkAway 对碳青霉烯耐药革兰阴性杆菌中头孢他啶-阿维巴坦与头孢洛扎-他唑巴坦药敏的评估

Evaluation of MicroScan WalkAway for Determination of Ceftazidime-Avibactam and Ceftolozane-Tazobactam Susceptibility in Carbapenem-Resistant Gram-Negative Bacilli.
PubMed 文献 · J Clin Microbiol · 2021-09-29

该研究评估了 MicroScan Neg 多重耐药 MIC 1(NMR1)面板检测碳青霉烯不敏感革兰阴性杆菌对头孢他啶-阿维巴坦和头孢洛扎-他唑巴坦的药敏性能,以微量肉汤稀释法为参考。共纳入 212 株临床分离株(139 株铜绿假单胞菌和 73 株产 KPC 肠杆菌目),来自巴西 71 家医院(2013 至 2020 年)。头孢他啶-阿维巴坦对铜绿假单胞菌和肠杆菌目的基本一致性均为 100%,分类一致性对肠杆菌目为 100%、对铜绿假单胞菌为 98.6%。

要点

  • NMR1 面板检测头孢他啶-阿维巴坦对产 KPC 肠杆菌目和碳青霉烯不敏感铜绿假单胞菌基本一致性和分类一致性高。
  • 头孢洛扎-他唑巴坦对铜绿假单胞菌的基本一致性为 98.6%,分类一致性为 96.4%。
  • 当头孢洛扎-他唑巴坦 MIC 为 4 mg/L 时,建议发出警示并用替代方法确认。

核心结论

  • NMR1 面板是检测产 KPC 肠杆菌目和碳青霉烯不敏感铜绿假单胞菌对头孢他啶-阿维巴坦药敏的一种选择。
  • 当头孢洛扎-他唑巴坦 MIC 为 4 mg/L 时,可创建警示并用替代方法确认结果。

研究设计与样本

  • 研究纳入 212 株临床显著的碳青霉烯不敏感革兰阴性杆菌,包括 139 株铜绿假单胞菌和 73 株产 KPC 肠杆菌目,来自 71 家巴西医院(2013 至 2020 年)。
  • 将 NMR1 面板测得的头孢他啶-阿维巴坦和头孢洛扎-他唑巴坦 MIC 与作为参考方法的微量肉汤稀释法(BMD)进行比较。
  • 评估基本一致性(EA)和分类一致性(CA);对铜绿假单胞菌采用误差率界限法计算药敏试验错误率。

主要发现

  • 头孢他啶-阿维巴坦对铜绿假单胞菌和肠杆菌目的基本一致性(总体和可评估)均为 100%。
  • 头孢他啶-阿维巴坦的分类一致性对肠杆菌目为 100%,对铜绿假单胞菌为 98.6%。
  • 头孢洛扎-他唑巴坦对铜绿假单胞菌的基本一致性(总体和可评估)分别为 98.6% 和 100%,分类一致性为 96.4%。
  • 对于头孢他啶-阿维巴坦,未发现极重大错误(VME),重大错误(ME)率为 4.2%(2/48)。
  • 对于头孢洛扎-他唑巴坦和铜绿假单胞菌,使用 CLSI 折点时微小错误(mE)为 11.4%,未发现 VME 或 ME;使用 EUCAST 折点时 VME 为 11.4%,未发现 ME。

关键数值与推荐

  • 头孢他啶-阿维巴坦对铜绿假单胞菌和肠杆菌目的基本一致性均为 100%。
  • 头孢他啶-阿维巴坦对肠杆菌目的分类一致性为 100%,对铜绿假单胞菌为 98.6%。
  • 头孢洛扎-他唑巴坦对铜绿假单胞菌的基本一致性为 98.6%(总体)和 100%(可评估),分类一致性为 96.4%。
  • 头孢他啶-阿维巴坦的 ME 率为 4.2%(2/48)。
  • 头孢洛扎-他唑巴坦对铜绿假单胞菌:CLSI 折点下 mE 为 11.4%,无 VME 或 ME;EUCAST 折点下 VME 为 11.4%,无 ME。

临床怎么用

  • NMR1 面板可用于检测产 KPC 肠杆菌目和碳青霉烯不敏感铜绿假单胞菌对头孢他啶-阿维巴坦的药敏。
  • 当头孢洛扎-他唑巴坦 MIC 为 4 mg/L 时,建议发出警示并用替代方法确认结果。

局限性

  • 研究仅纳入来自巴西 71 家医院的 212 株分离株,结果外推需谨慎。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

We evaluated the performance of ceftazidime-avibactam and ceftolozane-tazobactam MicroScan Neg multidrug-resistant MIC 1 (NMR1) panel for clinical carbapenem-nonsusceptible Gram-negative bacilli isolates. We evaluated 212 clinically significant carbapenem-nonsusceptible Gram-negative bacilli (139 Pseudomonas aeruginosa and 73 KPC-producing Enterobacterales) from 71 Brazilian hospitals (2013 to 2020). Ceftazidime-avibactam and ceftolozane-tazobactam MICs from the panel were compared with a broth microdilution (BMD) test as the reference method. Essential agreement (EA) and categorical agreement (CA) were assessed. For P. aeruginosa, antimicrobial susceptibility testing error rates were calculated using the error-rate bound method. Discrepancies were initially observed with 11 isolates; 4 resolved after retesting, 2 in favor of the NMR1 and 2 in favor of the BMD method. The ceftazidime-avibactam EA (overall and evaluable) was 100% for P. aeruginosa and Enterobacterales. The CA was 100% for Enterobacterales and 98.6% for P. aeruginosa. The ceftolozane-tazobactam EA was 98.6% and 100% (overall and evaluable, respectively), and the CA was 96.4% for P. aeruginosa. For ceftazidime/avibactam, no very major error (VME) was found, and the major error (ME) rate was 4.2% (2/48). For ceftolozane-tazobactam and P. aeruginosa, using the CLSI breakpoints, the minor error (mE) was 11.4%, and no VME or ME was found. While using EUCAST breakpoints, the VME was 11.4% with no ME. The mE becomes ME or VME in the absence of the intermediate category. All categorical errors were also within 1 log of MIC variation, and the adjusted error rate for CLSI/EUCAST was 0% (0/212). The NMR1 panel is an option to test ceftazidime-avibactam for KPC-producing Enterobacterales and carbapenem-nonsusceptible P. aeruginosa. When a MIC of 4 mg/liter for ceftolozane-tazobactam is obtained using this method, an alert could be created, and the results could be confirmed by an alternative method.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Ito CAS, Bail L, Arend LNVS, et al. Evaluation of MicroScan WalkAway for Determination of Ceftazidime-Avibactam and Ceftolozane-Tazobactam Susceptibility in Carbapenem-Resistant Gram-Negative Bacilli. J Clin Microbiol. 2021. PMID: 34586889. doi:10.1128/JCM.01536-21

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。