文献速递临床微生物质量管理与危急值

梯度扩散法(Etest)检测尿气球菌对氟喹诺酮类敏感性的验证

Validation of a Gradient Diffusion Method (Etest) for Testing of Antimicrobial Susceptibility of Aerococcus urinae to Fluoroquinolones.
PubMed 文献 · J Clin Microbiol · 2021-07-19

Etest与联合参考方法(BMD或AD)分类一致率环丙沙星95.2%、左氧氟沙星97.1%,无极重大误差。

要点

  • 尿气球菌对环丙沙星和左氧氟沙星敏感率分别为82.6%和81.6%。
  • Etest与联合参考方法分类一致率环丙沙星95.2%、左氧氟沙星97.1%。
  • Etest无极重大误差,重大误差率环丙沙星0.6%、左氧氟沙星1.2%。

核心结论

  • 在含5%羊血的阳离子调节MH琼脂上使用Etest检测尿气球菌对环丙沙星和左氧氟沙星的敏感性,与参考方法一致性良好,临床实验室可考虑采用。

研究设计与样本

  • 以CLSI M45-A3肉汤微量稀释法(BMD)为参考方法,EUCAST推荐的琼脂稀释法(AD)用于仲裁差异或处理技术问题。

主要发现

  • 207株中BMD检测37株(17.9%)拖尾、19株(9.2%)生长不足,改用AD检测。
  • 环丙沙星38株(18.4%)、左氧氟沙星13株(6.3%)Etest与BMD缺乏基本或分类一致,改用AD检测。
  • 采用联合参考方法(BMD或AD),尿气球菌对环丙沙星和左氧氟沙星的敏感率分别为82.6%和81.6%。
  • Etest与联合参考方法分类一致率环丙沙星95.2%、左氧氟沙星97.1%,无极重大误差。
  • 重大误差率和次要误差率环丙沙星分别为0.6%和4.3%,左氧氟沙星分别为1.2%和1.9%。

关键数值与推荐

  • Etest与联合参考方法分类一致率:环丙沙星95.2%,左氧氟沙星97.1%。
  • 重大误差率:环丙沙星0.6%,左氧氟沙星1.2%;次要误差率:环丙沙星4.3%,左氧氟沙星1.9%。
  • 敏感率:环丙沙星82.6%,左氧氟沙星81.6%(基于联合参考方法)。

临床怎么用

  • 临床实验室可采用含5%羊血的阳离子调节MH琼脂Etest对尿气球菌进行环丙沙星和左氧氟沙星敏感性检测。

局限性

  • BMD检测中部分菌株出现拖尾或生长不足,需AD仲裁。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Aerococcus urinae is a urinary pathogen with well-described resistance to fluoroquinolones. This study aimed to validate the gradient diffusion (GD) method (Etest) on cation-adjusted Mueller-Hinton agar with 5% sheep blood for testing the susceptibilities of Aerococcus urinae to the antimicrobial agents ciprofloxacin and levofloxacin and to compare the Etest to the broth microdilution (BMD) method from CLSI document M45-A3. Agar dilution (AD), as recommended by EUCAST, was used as an alternative reference method to arbitrate discrepancies or address technical issues. Aerococcus urinae isolates from urinary specimens were prospectively collected between June 2016 and December 2017 from six hospitals in Quebec, Canada, and identifications were confirmed using Vitek MS with the IVD 3.0 database. Of the 207 isolates tested using BMD, 37 (17.9%) showed trailing and 19 (9.2%) showed insufficient growth; these were tested using AD. Also, 38 isolates (18.4%) for ciprofloxacin and 13 isolates (6.3%) for levofloxacin showed a lack of essential or categorical agreement between the Etest and BMD and were also tested by AD. By use of a combined reference method (BMD or AD), the susceptibility rates of Aerococcus urinae were 82.6% and 81.6% for ciprofloxacin and levofloxacin, respectively. Categorical agreement between GD and the combined reference methods was 95.2% for ciprofloxacin and 97.1% for levofloxacin, with no very major error identified. Major and minor error rates were 0.6% and 4.3% for ciprofloxacin and 1.2% and 1.9% for levofloxacin. Overall, antimicrobial susceptibility testing (AST) using the Etest on sheep blood agar showed good agreement with the reference methods and can be considered by clinical laboratories wishing to perform AST on Aerococcus urinae isolates.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Roy FE, Berteau T, Bestman-Smith J, et al. Validation of a Gradient Diffusion Method (Etest) for Testing of Antimicrobial Susceptibility of Aerococcus urinae to Fluoroquinolones. J Clin Microbiol. 2021. PMID: 34076476. doi:10.1128/JCM.00259-21

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。