文献速递临床微生物质量管理与危急值

全自动纸片扩散法药敏试验:Copan WASP Colibri 联合 Radian 的性能评估

Performance of Fully Automated Antimicrobial Disk Diffusion Susceptibility Testing Using Copan WASP Colibri Coupled to the Radian In-Line Carousel and Expert System.
PubMed 文献 · J Clin Microbiol · 2021-08-18

本研究在常规实验室条件下比较 Vitek 2 与 Colibri 联合 Radian 全自动纸片扩散法药敏试验的分类一致性。共纳入 718 株菌(含 43 株碳青霉烯酶肠杆菌),总体分类一致率 99.3%(4,350/4,380;95%CI 99%~99.5%),Colibri 联合 Radian 的准确性等于或优于 Vitek 2。

要点

  • Colibri 联合 Radian 提供全自动纸片扩散法药敏方案,准确性等于或优于 Vitek 2。
  • Vitek 2 对铜绿假单胞菌极重大错误的主要原因为异质性耐药群体(62%,13/21)。
  • Vitek 2 漏检 1 株 OXA-48 样和 1 株 OXA-181 样碳青霉烯酶肠杆菌,即使严格应用 EUCAST 筛选折点。

核心结论

  • Colibri 联合 Radian 的全自动纸片扩散法药敏试验准确性等于或优于 Vitek 2 系统。

研究设计与样本

  • 菌株包括 249 株肠杆菌目、198 株铜绿假单胞菌、107 株金黄色葡萄球菌、78 株凝固酶阴性葡萄球菌、38 株粪肠球菌和 5 株屎肠球菌。
  • 在常规实验室条件下评估 Vitek 2 与 Colibri 联合 Radian 在分类水平上的一致性。

主要发现

  • 总体分类一致率为 99.3%(4,350/4,380;95%CI 99%~99.5%)。
  • 肠杆菌目分类一致率 98.6%(2,147/2,178;95%CI 98.0%~99.0%)。
  • 铜绿假单胞菌分类一致率 99.4%(1,839/1,850;95%CI 98.9%~99.7%)。
  • 葡萄球菌属分类一致率 99.4%(342/344;95%CI 97.9%~99.8%)。
  • Vitek 2 对铜绿假单胞菌极重大错误的主要原因为异质性耐药群体(62%,13/21)。

关键数值与推荐

  • 肠杆菌目分类一致率 98.6%(2,147/2,178;95%CI 98.0%~99.0%)。
  • 铜绿假单胞菌分类一致率 99.4%(1,839/1,850;95%CI 98.9%~99.7%)。
  • 葡萄球菌属分类一致率 99.4%(342/344;95%CI 97.9%~99.8%)。
  • 肠球菌属分类一致率 99.4%(342/344;95%CI 97.9%~99.8%)。
  • 总体分类一致率 99.3%(4,350/4,380;95%CI 99%~99.5%)。

临床怎么用

  • Colibri 联合 Radian 可作为全自动纸片扩散法药敏试验方案,准确性等于或优于 Vitek 2。
  • 对铜绿假单胞菌异质性耐药及碳青霉烯酶肠杆菌,需注意 Vitek 2 可能漏检。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The purpose of the present study was to assess the agreement at the categorical level between the Vitek 2 system and the Colibri coupled to the Radian under real routine laboratory conditions. The 675 nonduplicate clinical strains included in this study (249 Enterobacterales isolates, 198 Pseudomonas aeruginosa, 107 Staphylococcus aureus, 78 coagulase-negative staphylococci, 38 Enterococcus faecalis, and 5 Enterococcus faecium) were isolated from nonconsecutive clinical samples referred to our laboratory between June and November 2020. In addition, 43 carbapenemase-producing Enterobacterales (CPE) formerly identified and stored in our laboratory were added to the panel, for a total of 718 strains. The overall categorical agreements between the two compared methods were 99.3% (4,350/4,380; 95% CI 99% to 99.5%); 98.6% (2,147/2,178; 95% CI 98.0% to 99.0%); 99.4% (1,839/1,850; 95% CI 98.9% to 99.7%); and 99.4% (342/344; 95% CI 97.9% to 99.8%) for Enterobacterales, P. aeruginosa, Staphylococcus spp., and Enterococcus spp., respectively. The most important cause of the very major errors encountered on the Vitek 2 for P. aeruginosa (62%, 13/21) was related to the presence of heteroresistant populations. Among the 43 CPE included in this study, one OXA-48-like, and one OXA-181-like were missed by the Vitek 2, even by rigorously applying the CPE screening cutoffs defined by EUCAST. The Colibri coupled to the Radian provide a fully automated solution for antimicrobial disk diffusion susceptibility testing with an accuracy that is equal to or better than that of the Vitek 2 system.

来源:J Clin Microbiol,查看原文。版权归原作者所有。

引用

  1. Cherkaoui A, Renzi G, Vuilleumier N, et al. Performance of Fully Automated Antimicrobial Disk Diffusion Susceptibility Testing Using Copan WASP Colibri Coupled to the Radian In-Line Carousel and Expert System. J Clin Microbiol. 2021. PMID: 34160274. doi:10.1128/JCM.00777-21

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。