文献速递临床免疫质量管理与危急值

三种自动化降钙素原免疫分析法在医学决策点的一致性比较

Concordance of Three Automated Procalcitonin Immunoassays at Medical Decision Points.
PubMed 文献 · Ann Lab Med · 2021-01-01

该研究纳入119份降钙素原浓度<5.00 μg/L的血清样本,以Kryptor为参考方法,比较Atellica IM 1600和Cobas e801在0.25和0.50 μg/L两个抗生素管理决策点的一致性。两种方法相关性高(ρ=0.97和0.99),但均存在负偏倚;尽管线性加权κ值达0.90和0.92,0.25 μg/L处差异更明显,故不建议重复检测中交替使用不同PCT免疫分析法。

要点

  • Atellica IM 1600和Cobas e801与Kryptor相关性高,但均存在负偏倚。
  • 两法在0.25和0.50 μg/L决策点一致性优秀,但0.25 μg/L处差异更明显。
  • 重复检测中不建议交替使用不同PCT免疫分析法,标准化前不宜直接比较结果。

核心结论

  • Atellica IM 1600和Cobas e801与Kryptor在0.25和0.50 μg/L决策点一致性优秀,但存在负偏倚和差异,不建议重复检测中交替使用不同PCT免疫分析法。

研究设计与样本

  • 研究纳入119份PCT浓度<5.00 μg/L的血清样本,以Kryptor为参考方法,用Spearman秩相关、回归分析和两个医学决策点(0.25和0.50 μg/L)的一致性比较Atellica IM 1600和Cobas e801。

主要发现

  • Atellica IM 1600和Cobas e801与Kryptor的相关系数ρ分别为0.97和0.99。
  • Atellica IM 1600的斜率/截距为0.75/-0.00,Cobas e801为0.88/-0.01,均显示负偏倚。
  • 在0.25和0.50 μg/L决策点,Atellica IM 1600和Cobas e801的线性加权κ值分别为0.90和0.92,但差异在0.25 μg/L处更明显。

关键数值与推荐

  • 医学决策点为0.25和0.50 μg/L。
  • Atellica IM 1600和Cobas e801与Kryptor的线性加权κ值分别为0.90和0.92。

临床怎么用

  • 临床解读PCT结果时应考虑不同免疫分析法间的偏倚,重复检测中避免交替使用不同方法,标准化前不宜直接比较不同方法的结果。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Procalcitonin (PCT) is a useful bacterial infection biomarker with the potential for guiding antibiotic therapy. We evaluated the concordance of three automated PCT immunoassays: Kryptor (BRAHMS GmbH, Hennigsdorf, Germany), Atellica IM 1600 (Siemens Healthcare Diagnostics, Munich, Germany), and Cobas e801 (Roche Diagnostics, Mannheim, Germany). In 119 serum samples with a PCT concentration <5.00 μg/L, Kryptor (reference assay) was compared with the other two immunoassays by Spearman's rank correlation, regression analysis, and concordance at two antibiotic stewardship medical decision points: 0.25 and 0.50 μg/L. The Atellica IM 1600 and Cobas e801 results showed high correlations with those of Kryptor, with correlation coefficient (ρ) values of 0.97 and 0.99, respectively. However, negative biases were observed in both immunoassays (slope/y-intercept: 0.75/-0.00 for Atellica IM 1600; 0.88/-0.01 for Cobas e801). Atellica IM 1600 and Cobas e801 demonstrated excellent concordance with Kryptor at both medical decision points, with linearly weighted κ values of 0.90 and 0.92, respectively, despite discrepancies, which were more prominent at the 0.25 μg/L medical decision point. Based on these biases and discrepancies, the alternate use of different PCT immunoassays in repeat examinations is inadvisable. Standardization is required before comparing the results of different PCT immunoassays.

来源:Ann Lab Med,查看原文。版权归原作者所有。

引用

  1. Cho HW, Kim SH, Cho Y, et al. Concordance of Three Automated Procalcitonin Immunoassays at Medical Decision Points. Ann Lab Med. 2021. PMID: 33536362. doi:10.3343/alm.2021.41.4.419

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。