患者实时质量控制算法对不同类型分析误差的性能评估
该研究基于复旦大学附属中山医院2019年门诊血清钠、氯、丙氨酸氨基转移酶和肌酐共906,552个检测结果,比较六种患者实时质量控制(PBRTQC)算法对恒定误差、比例误差和随机误差的检测性能。结果显示移动均值、移动中位数、指数加权移动均值和移动四分位数对恒定误差和比例误差检测效果相似但对随机误差无效;移动阳性患者和与移动标准差可检测血清钠和氯的随机误差但对恒定误差和比例误差检测效果差。研究强调选择质控算法前需评估分析物潜在误差来源及对应分析误差类型。
要点
- 移动均值、移动中位数、指数加权移动均值和移动四分位数可有效检测恒定误差和比例误差。
- 上述四种算法对随机误差检测效果不佳。
- 移动阳性患者和与移动标准差可检测血清钠和氯的随机误差,但对恒定误差和比例误差检测效果差。
核心结论
- 选择PBRTQC算法前应评估特定分析物的潜在误差来源及对应的分析误差类型。
研究设计与样本
- 比较研究,使用复旦大学附属中山医院检验科2019年门诊血清钠、氯、丙氨酸氨基转移酶和肌酐共906,552个检测结果,评估六种PBRTQC算法对三种分析误差类型的性能。
主要发现
- 移动均值、移动中位数、指数加权移动均值和移动四分位数对恒定误差和比例误差检测效果相似,但对随机误差无效。
- 移动阳性患者和与移动标准差可检测血清钠和氯的随机误差,但对恒定误差和比例误差检测效果差。
临床怎么用
- 实施PBRTQC前应明确分析物的潜在误差来源及对应分析误差类型,据此选择合适算法。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Patient-based real-time quality control (PBRTQC) has gained attention because of its potential to detect analytical errors in situations wherein internal quality control is less effective. Multiple PBRTQC algorithms have been proposed. However, there is a lack of comprehensive comparison of the performance of PBRTQC algorithms on different types of analytical errors. Thus, a comparative study was conducted. METHODS: The performance of six different PBRTQC algorithms was evaluated on three types of analytical errors using 906,552 test results for outpatient serum sodium, chloride, alanine aminotransferase, and creatinine at the Department of Laboratory Medicine at Zhongshan Hospital, Fudan University in 2019. The performance results were compared and assessed. RESULTS: The moving average, moving median, exponentially weighted moving average, and moving quartiles performed similarly for effectively detecting constant errors (CE) and proportional errors (PE) but not random errors (RE). The moving sum of positive patients and moving standard deviation could detect RE for serum sodium and chlorides but performed poorly on detecting the CE and PE. CONCLUSIONS: This study demonstrated the importance of assessing the potential source of error of a particular analyte and the corresponding type of analytical error before choosing a quality control algorithm for implementation.
引用
- Duan X, Wang B, Zhu J, et al. Assessment of patient-based real-time quality control algorithm performance on different types of analytical error. Clin Chim Acta. 2020. PMID: 33127347. doi:10.1016/j.cca.2020.10.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。