液相色谱-串联质谱法快速可靠定量胰高血糖素的方法开发与验证
本研究开发并验证了一种基于液相色谱-串联质谱(LC-MS/MS)的胰高血糖素定量方法,采用混合模式阴离子交换固相萃取从血浆中提取胰高血糖素,并评估了样本稳定性。该方法检测不精密度低于4%,回收率在95-103%之间,定量限为8.75 pg/mL,总分析CV为2.91%,样本在4°C下稳定不足4小时。与两种免疫分析法比较,Diasource® RIA与本方法不一致,Mercodia® ELISA一致性更接近,外部质控样本也证实了这一点。结论认为该方法快速、特异,适合常规检测,并有望成为标准化方法。
要点
- LC-MS/MS法检测胰高血糖素不精密度低于4%,回收率95-103%,定量限8.75 pg/mL。
- 样本在4°C下稳定不足4小时,需注意前处理时限。
- 与Diasource® RIA不一致,与Mercodia® ELISA一致性更接近。
核心结论
- 开发并验证了一种快速、特异的LC-MS/MS法用于胰高血糖素定量,适合常规检测,并有望成为标准化方法。
研究设计与样本
- 采用混合模式阴离子交换固相萃取从血浆中提取胰高血糖素,评估样本在K2-EDTA和P800管中不同温度下的稳定性,与两种免疫分析法比较,遵循FDA和EMA指南进行验证,并利用外部质控计划与其他实验室比较。
主要发现
- 检测不精密度低于4%,回收率在95-103%之间,定量限为8.75 pg/mL,总分析CV为2.91%。
- 样本在4°C下稳定不足4小时。
- Diasource® RIA与本方法不一致,Mercodia® ELISA一致性更接近,外部质控样本也证实了这一点。
关键数值与推荐
- 检测不精密度低于4%,回收率95-103%,定量限8.75 pg/mL,总分析CV为2.91%。
- 样本在4°C下稳定不足4小时。
临床怎么用
- 该方法可用于常规胰高血糖素定量,但需注意样本在4°C下稳定不足4小时,应尽快处理。
- 与Mercodia® ELISA一致性更接近,与Diasource® RIA不一致,临床解读时需考虑方法差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The quantitation of glucagon remains a challenging immunoassay, mainly due to cross-reactivity. A sensitive, rapid and specific intact glucagon method is therefore necessary for quality routine analysis. A tandem mass spectrometry method to fulfill this objective is described in this work. METHODS: Glucagon was extracted from plasma employing a mixed-mode anion exchange solid-phase extraction. Sample stability was assessed in K2-EDTA and P800 tubes at different temperatures. We compared our method to two different immunoassays. FDA and EMA guidelines were followed for validation. An external quality control program served for comparison with other laboratories. RESULTS: Assay imprecision was below 4%. Recoveries were within 95-103%. LoQ was 8.75 pg/mL. Total analytical CV was 2.91%. Samples were found stable at 4 °C for less than 4 h. Diasource® RIA disagreed with our method. Mercodia® ELISA provided a closer agreement, also proven by external quality control samples. CONCLUSIONS: A rapid and specific LC-MS/MS method for glucagon quantitation has been developed, validated and is suitable to routine care. The simplicity and the good performances in terms of time and specificity, could open the possibility to establish a standardized method for glucagon.