PhenoMatrix算法对ChromID Strepto B培养基上B族链球菌数字图像分析的检测性能
该研究评估PhenoMatrix(PM)显色检测模块(CDM)数字成像软件在WASP自动处理器上从ChromID Strepto B显色培养基检测B族链球菌(GBS)的临床性能,并与人工判读数字图像及分子检测比较。以复合参考标准判断,PM CDM检测GBS的敏感性为100%,未遗漏48小时孵育内的真阳性分离株;ChromID联合PM CDM的敏感性(95.5%)与核酸扩增检测(96.8%)相近,但特异性(63.0%)低于ChromID单独(95.4%)。
要点
- PM CDM对ChromID培养基检测GBS的敏感性达100%,未遗漏48小时内的真阳性。
- ChromID联合PM CDM敏感性95.5%,与NAAT的96.8%相近。
- ChromID联合PM CDM特异性63.0%,低于ChromID单独培养的95.4%。
核心结论
- PM CDM检测ChromID培养基上GBS的敏感性为100%,未遗漏48小时孵育内的真阳性分离株。
- ChromID联合PM CDM的敏感性(95.5%)与分子检测(96.8%)相近,但特异性(63.0%)较低。
研究设计与样本
- 研究评估PhenoMatrix显色检测模块数字成像软件在WASP自动处理器上从ChromID Strepto B培养基检测GBS的性能。
- PM CDM结果与人工判读数字图像及GBS分子检测进行比较。
主要发现
- 以核酸扩增检测(NAAT)为参考,ChromID单独培养48小时的敏感性和特异性分别为84.5%和94.7%。
- 以复合参考标准判断,PM CDM检测GBS的敏感性为100%,未遗漏48小时孵育内的真阳性分离株。
- 评估三种方法检测GBS:NAAT、ChromID联合PM CDM(48小时)和ChromID单独(48小时)的敏感性分别为96.8%、95.5%和90.3%。
- NAAT、ChromID联合PM CDM和ChromID单独的特异性分别为97.7%、63.0%和95.4%。
- 算法从未将人工判读阳性的培养判为阴性,并额外识别出8份人工数字图像判读遗漏的真阳性标本。
临床怎么用
- PM CDM数字成像软件可辅助自动检测ChromID培养基上的GBS,提高敏感性并减少人工判读遗漏。
- 需注意ChromID联合PM CDM时特异性较低(63.0%),可能增加假阳性结果,应结合临床判断。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Group B Streptococcus (GBS) can be found to colonize about 25% of all healthy, adult women and is the leading infectious cause of early neonatal morbidity and mortality in the United States. This study evaluated the clinical performance of PhenoMatrix (PM) chromogenic detection module (CDM) digital imaging software in detection of GBS from LIM broth plated on ChromID Strepto B chromogenic medium (ChromID) using the WASP automated processor. The performance of the PM CDM was compared to manual culture review of the digital images and molecular detection of GBS. ChromID alone had a sensitivity and specificity of 84.5% and 94.7%, respectively, after 48 h compared to nucleic acid amplification testing (NAAT). Compared to the composite reference for positivity, when PM CDM was used to detect GBS from ChromID, the sensitivity was 100%, with no true-positive GBS isolates missed by 48 h of incubation. Overall, evaluating all three methods for the detection of GBS, the sensitivities of NAAT, ChromID plus PM CDM at 48 h, and ChromID alone at 48 h were 96.8%, 95.5%, and 90.3%, respectively. The specificities of NAAT, ChromID plus PM CDM, and ChromID alone were 97.7%, 63.0%, and 95.4%, respectively. The sensitivity of ChromID in combination with the PM CDM was similar to the sensitivity of molecular detection. Further, the algorithm never called a culture negative that was determined to be positive by manual reading, and it identified an additional eight true positive specimens that were missed by manual digital image culture reading.
引用
- Baker J, Timm K, Faron M, et al. Digital Image Analysis for the Detection of Group B Streptococcus from ChromID Strepto B Medium Using PhenoMatrix Algorithms. J Clin Microbiol. 2020. PMID: 33087433. doi:10.1128/JCM.01902-19