V8毛细管电泳系统CDT检测性能评估及与IFCC认可HPLC法和N-Latex免疫比浊法的比较
本研究评估了Helena Biosciences V8电泳分析仪检测碳水化合物缺乏转铁蛋白(CDT)的分析性能,并在231例患者样本中与IFCC认可的高效液相色谱(HPLC)法和N-Latex CDT免疫比浊法进行比较。V8法与两法的相关性r²≥0.97,分类CDT升高者的Kappa≥0.8;以HPLC分类为参考,V8法的ROC曲线下面积为0.996(95%CI 0.99-1.00)。结论认为V8法稳健、可靠且有效。
要点
- V8法总不精密度为5.1%–24.3%,浓度≥1.62%的样本不精密度≤13.1%。
- V8法与HPLC和N-Latex法的相关性r²≥0.97,分类CDT升高者的Kappa≥0.8。
- 以HPLC分类为参考,V8法的ROC曲线下面积为0.996(95%CI 0.99-1.00)。
核心结论
- V8法CDT检测与HPLC和N-Latex法高度相关,分类CDT升高者的一致性极好。
- V8法CDT分析是检测血清CDT浓度的稳健、可靠且有效的方法。
研究设计与样本
- 在兩個独立地点研究V8法的分析性能,包括线性和不精密度。
- 在231例患者样本中比较V8、HPLC和N-Latex法的相关性、偏倚和一致性。
主要发现
- V8法总不精密度范围为5.1%–24.3%,浓度≥1.62%的样本不精密度≤13.1%。
- 所有方法比较的相关性r²≥0.97。
- V8法相对于HPLC法的偏倚为-0.83(-22.24%),相对于N-Latex法的偏倚为-0.40(-12.26%)。
- 各方法分类CDT升高者的一致性极好,Kappa评分≥0.8。
- 以HPLC分类为参考,N-Latex法和V8法的ROC曲线下面积分别为0.981(95%CI 0.97-0.99)和0.996(95%CI 0.99-1.00)。
关键数值与推荐
- 临床cut-off值为≥1.62%。
- V8法与HPLC法偏倚-0.83(-22.24%),与N-Latex法偏倚-0.40(-12.26%)。
- 分类CDT升高者的Kappa评分≥0.8。
- V8法ROC曲线下面积0.996(95%CI 0.99-1.00),N-Latex法0.981(95%CI 0.97-0.99)。
临床怎么用
- V8法可用于血清CDT检测,与HPLC和N-Latex法分类CDT升高者的一致性极好。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Carbohydrate-deficient transferrin (CDT) measurements are commonly used for the identification and follow-up of individuals suspected of chronic alcohol abuse. This study describes the analytical characteristics of the CDT assay on the Helena Biosciences V8 electrophoresis analyzer and compares its diagnostic performance to the International Federation of Clinical Chemistry and Laboratory Medicine approved high performance liquid chromatography (HPLC) method and the N-Latex CDT immunonephelometric assay. METHODS: The analytical performance of the V8 assay, including the linearity and the imprecision, was studied at two separate locations. Method comparison analysis was performed by studying the correlation, bias and agreement between the V8, HPLC and the N-Latex assays in 231 patient samples. RESULTS: The total imprecision ranged between 5.1 and 24.3% and was ≤13.1% for samples with concentrations above the clinical cut-off value (≥1.62%). The method comparisons revealed excellent correlations with r2≥0.97 for all comparisons. Measurements on the V8 showed a bias of -0.83 (-22.24%) and -0.40 (-12.26%) with the HPLC and N-Latex assays, respectively. The assays showed excellent agreements (Kappa scores ≥ 0.8) in classifying subjects with elevated CDT values. Receiver operating characteristic (ROC)-curve analysis, using the HPLC classification as reference, revealed areas under the ROC-curves of 0.981 (95% CI, 0.97-0.99) and 0.996 (0.99-1.00) for the N-Latex and V8 assays, respectively. CONCLUSIONS: CDT measurements on the V8 assay are highly correlated with both the HPLC and the N-Latex assay and show excellent agreement in classifying subjects with elevated CDT values. Overall, the V8 CDT analysis is a robust, reliable and effective method to measure CDT concentrations in serum samples.
引用
- Jacobs LHJ, Te Stroet RM, Demir AY, et al. Evaluation of carbohydrate-deficient transferrin measurements on the V8 capillary electrophoresis system and comparison with the IFCC approved HPLC reference method and N-Latex immunonephelometric assay. Clin Chem Lab Med. 2020. PMID: 33048832. doi:10.1515/cclm-2020-0545
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。