自动化临床化学与微生物实验室常规检测中微生物污染风险评估
该研究在自动化临床化学和微生物实验室中,用荧光标记和MS2病毒示踪评估常规检测时的污染风险。结果显示仪器污染概率低,但污染标本容器外表面可致环境表面污染,且个人防护装备使用和手卫生依从性不佳。
要点
- 自动化仪器在处理严重污染标本时污染风险低。
- 污染容器外表面可导致手套、电脑配件和标本架等环境表面污染。
- 个人防护装备使用不规范和手卫生依从性低增加污染风险。
核心结论
- 自动化化学和微生物设备分析严重污染标本时仪器污染可能性低,但处理污染容器可污染环境表面,且个人防护装备依从性低加剧风险。
研究设计与样本
- 在自动化临床化学和微生物实验室进行3次实验,用荧光标记和MS2病毒示踪评估污染。
主要发现
- 165个实验室表面和人员中49个(30%)观察到荧光,93台全实验室自动化仪器中21台(23%)有荧光。
- 荧光最常转移至手套(31/40)、电脑配件(9/18)和标本加载架(12/12)。
- 123个拭子区域均未检出MS2。
- 化学和微生物实验室个人防护装备使用不当发生率分别为每 staff 每小时0.36和0.15次。
- 手卫生依从性为132名评估人员中61名(46%)。
临床怎么用
- 处理污染标本容器时需加强环境表面清洁和个人防护装备规范使用。
- 提高手卫生依从性以减少实验室环境污染风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Every clinical specimen is potentially infectious, but data regarding risk for contamination of the laboratory environment during routine testing are scarce. We assessed contamination during routine sample analysis in automated clinical chemistry and microbiology laboratories. METHODS: A fluorescent marker was applied to specimen container exteriors to assess the impact of gross contamination. Nonpathogenic MS2 virus was added to remnant blood, urine, and ESwab matrices as a biomarker of cross-contamination. Samples were processed and analyzed using Roche Cobas 8100 and ISE, c502, e602, and c702 modules (blood) and BD Kiestra total laboratory automation (blood, urine, ESwabs) over 3 experiments. Fluorescence transfer to laboratory surfaces and personnel was visualized using ultraviolet light. Surfaces were swabbed and assessed for MS2 cross-contamination by RT-PCR. Adherence to standard precautions by laboratory staff was assessed by observation. RESULTS: Fluorescence was observed on 49 of 165 (30%) laboratory surfaces and personnel and 21 of 93 (23%) total laboratory automation instruments. Fluorescence transferred most frequently to gloves (31/40), computer accessories (9/18), and specimen loading racks (12/12). None of 123 areas swabbed were positive for MS2. Improper personal protective equipment use occurred at a rate of 0.36 and 0.15 events per staff per hour in the chemistry and microbiology laboratories, respectively. Hand-washing compliance was observed for 61 of 132 (46%) staff members evaluated. CONCLUSIONS: Analysis of grossly contaminated specimens on automated chemistry and microbiology equipment elicits a low likelihood of instrument contamination. However, handling contaminated specimen containers can result in contamination of environmental laboratory surfaces, representing a source of risk that is heightened by low adherence to appropriate personal protective equipment.
引用
- Farnsworth CW, Wallace MA, Liu A, et al. Evaluation of the Risk of Laboratory Microbial Contamination during Routine Testing in Automated Clinical Chemistry and Microbiology Laboratories. Clin Chem. 2020. PMID: 32870987. doi:10.1093/clinchem/hvaa128
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。