热灭活对ELISA检测SARS-CoV-2 IgM和IgG抗体的影响
该研究收集62例COVID-19患者和18例健康对照的血清,比较热灭活前后ELISA检测SARS-CoV-2 IgM和IgG抗体的水平。结果显示灭活后抗体指标显著升高,但阳性率不变,灭活前后IgM和IgG的符合率均很高,结论是56°C 30分钟热灭活不降低诊断效能,推荐用于减少实验室人员感染风险。
要点
- 热灭活后SARS-CoV-2 IgM和IgG抗体水平显著高于新鲜血清。
- 灭活前后所有参与者和COVID-19患者的抗体阳性率均无变化。
- 推荐56°C 30分钟热灭活血清以降低实验室人员病毒污染风险。
核心结论
- 热灭活不降低SARS-CoV-2 IgM或IgG抗体的诊断效能,推荐56°C 30分钟灭活血清。
研究设计与样本
- 收集2020年2月27日至3月6日武汉汉口医院62例COVID-19患者和18例健康对照的血清样本。
- 比较灭活前后ELISA检测的SARS-CoV-2 IgM和IgG抗体水平。
主要发现
- 灭活后抗体指标显著高于新鲜血清,但所有参与者和COVID-19患者的阳性率不变。
- IgM灭活前后阳性符合率100.00%(55/55),阴性符合率96.00%(24/25),总符合率98.75%(79/80),κ=0.971,P<0.001。
- IgG灭活前后阳性符合率98.21%(55/56),阴性符合率91.67%(22/24),总符合率98.75%(79/80),κ=0.910,P<0.001。
关键数值与推荐
- IgM抗体灭活前后一致性κ=0.971,P<0.001。
- IgG抗体灭活前后一致性κ=0.910,P<0.001。
- 推荐血清经56°C 30分钟热灭活。
临床怎么用
- 在ELISA检测SARS-CoV-2 IgM和IgG抗体时,可采用56°C 30分钟热灭活血清以减少实验室人员病毒暴露风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: To establish a safe and accurate method for detecting SARS-CoV-2 IgM and IgG, we assessed the impact of sera after heat-inactivation on the SARS-CoV-2 IgM and IgG levels measured by ELISA-immunoassay. METHODS: The serum samples of 62 patients with COVID-19 and 18 healthy controls were collected in Hankou's Hospital of Wuhan from February 27 to March 6, 2020. Before and after the samples were inactivated, the levels of IgM and IgG antibodies were measured. RESULTS: The indexes of antibodies after inactivated were significantly higher than those in fresh sera, while the positive rates in all participants or in patients with COVID-19 did not change. The positive coincidence rate, negative coincidence rate and total coincidence rate of IgM antibodies before and after inactivation were 100.00% (55/55), 96.00% (24/25) and 98.75% (79/80), respectively (κ = 0.971, P < 0.001), while those for IgG antibodies were 98.21% (55/56), 91.67% (22/24) and 98.75% (79/80) respectively (κ = 0.910, P < 0.001). These results showed a good consistency. CONCLUSIONS: Heating-activation does not decrease the diagnostic efficacy of SARS-CoV-2 IgM or IgG antibodies. Sera inactivated by heating at 56 °C for 30 min should be recommended to minimize the risk of virus contamination of laboratory staff.
引用
- Hu X, Zhang R, An T, et al. Impact of heat-inactivation on the detection of SARS-CoV-2 IgM and IgG antibody by ELISA. Clin Chim Acta. 2020. PMID: 32569631. doi:10.1016/j.cca.2020.06.032
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。