糖基化可能影响性激素结合球蛋白的测量
研究用不同SHBG抗体配对的双位点ELISA比较测量差异,发现一对单克隆抗体(7H9)与基础包被抗体(11F11)配对时出现结果不一致。唾液酸酶处理使不一致样本的SHBG水平升高,提示糖基化影响抗体识别,而非异嗜性抗体所致。
要点
- SHBG双位点ELISA结果不一致可能与糖基化影响抗体识别有关。
- 7H9单克隆抗体识别SHBG可被覆盖O-连接糖基化位点的肽段干扰。
- 免疫分析不一致时需考虑糖基化因素而非仅归因于异嗜性抗体。
核心结论
- SHBG免疫分析结果不一致可能归因于糖基化影响抗体识别,而非异嗜性抗体或其他循环抗体。
研究设计与样本
- 使用多种自备SHBG抗体与一致对照配对,研究双位点ELISA测量不一致现象。
主要发现
- 以11F11为基础包被、7H9为顶层单克隆抗体时观察到不一致,多克隆SHBG抗体与11F11配对时亦出现不一致。
- 唾液酸酶处理后,不一致样本及纯化SHBG用7H9测得水平升高,而一致样本无变化。
- 唾液酸酶处理后,不一致样本用其他单克隆抗体配对测得SHBG无变化,而一致样本用相同配对测得SHBG水平下降。
- 覆盖O-连接糖基化位点的合成肽段可破坏7H9单克隆抗体对SHBG的识别。
临床怎么用
- 检验科遇到SHBG免疫分析结果与临床不符时,应考虑到糖基化影响抗体识别的可能性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
A discordance between sex hormone-binding globulin (SHBG) measurements by 2-site ELISAs was investigated using pairings of various "in house" SHBG antibodies together with a concordant control. The 2-site monoclonal ELISAs used the same base coat (11F11) and discordance was observed with one top coat monoclonal antibody (7H9) and also when a polyclonal SHBG antibody was paired with the basecoat antibody (11F11). Sialidase treatment of the discordant sample and purified SHBG revealed increased levels using 7H9 whereas there was no change in SHBG in the concordant sample. Conversely, following sialidase treatment, the discordant sample showed no change in SHBG measured using the other monoclonal antibody pairings whereas the SHBG levels in the concordant sample declined following sialidase using the same monoclonal antibody pairings. This implicated glycosylation as a factor in antibody recognition and synthetic peptides spanning the two N-linked and one O-linked glycosylation regions showed that SHBG recognition by monoclonal antibody 7H9 could be disrupted by a peptide spanning the O-linked glycosylation site. Hence rather than immunoassay discordance being attributed to heterophile antibodies or other circulating antibodies here it can be likely attributed to glycosylation affecting antibody recognition and hence the measurement of SHBG.
引用
- Lewis JG, Elder PA. Glycosylation may influence sex hormone-binding globulin measurements. Clin Chim Acta. 2020. PMID: 32531254. doi:10.1016/j.cca.2020.06.016