Panther Fusion系统开放功能用于流感A病毒亚型实验室自建检测
研究在Panther Fusion平台上开发了LDT-FAST与exWHO-FAST两种流感A病毒亚型分型的实验室自建检测,并评估其与FFABR试剂并行检测的性能。110份流感A病毒阳性样本中,104份既往已分型(54份H3、46份09H1、4份fsH1),两种LDT均正确分型;未分型样本中3份被两种LDT分型为H3、2份仅被LDT-FAST分型为H3。LDT-FAST的检出限与FFABR相近,exWHO-FAST对两株H3N2的检出限较高,与引物/探针同源性分析一致。
要点
- 两种LDT可在Panther Fusion上与FFABR试剂并行检测,实现流感A病毒亚型分型。
- 104份既往分型样本(54份H3、46份09H1、4份fsH1)均被两种LDT正确分型。
- LDT-FAST检出限与FFABR相近,exWHO-FAST对两株H3N2检出限较高。
核心结论
- 两种流感A病毒分型LDT均可作为Panther Fusion呼吸道检测菜单的有价值补充。
- 系统全自动化且可拆分洗脱液,支持IVD与LDT同时检测。
研究设计与样本
- 研究在Panther Fusion仪器上开发LDT-FAST与exWHO-FAST两种流感A病毒分型LDT,并与FFABR试剂并行检测。
- 性能评估使用110份流感A病毒阳性样本,其中104份既往已分型。
- 检出限分析使用5株流感A病毒株。
主要发现
- 104份既往分型样本中54份为H3、46份为09H1、4份为fsH1,两种LDT均正确分型。
- 未分型样本中3份被两种LDT分型为H3,2份仅被LDT-FAST分型为H3。
- 两种LDT均未分型的1份样本经FFABR复测转为阴性。
关键数值与推荐
- 既往分型样本构成:H3 54份、09H1 46份、fsH1 4份。
- LDT-FAST检出限与FFABR相近;exWHO-FAST对两株H3N2检出限较高,与引物/探针同源性分析一致。
临床怎么用
- 可在Panther Fusion上以LDT与IVD并行方式开展流感A病毒亚型分型,补充呼吸道检测菜单。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Nucleic acid amplification tests, such as PCR, are the method of choice for respiratory virus testing, due to their superior diagnostic accuracy and fast turnaround time. The Panther Fusion (Fusion; Hologic) system has an array of highly sensitive in vitro diagnostic (IVD) real-time PCR assays for respiratory viruses, including an assay for influenza A (FluA) virus, influenza B (FluB) virus, and respiratory syncytial virus (RSV) (FFABR assay). The Fusion system has Open Access functionality to perform laboratory-developed tests (LDTs) alongside IVD assays. We developed two LDTs for FluA virus strain typing on the Panther Fusion instrument, enabling side-by-side testing with the FFABR assay. The LDT-FAST assay uses proprietary primers and probes designed by Hologic for the Prodesse ProFAST+ (PFAST) assay. The exWHO-FAST assay is an expanded redesign of the WHO-recommended reverse transcriptase PCRs (RT-PCRs). To evaluate the performance of these two LDTs, 110 FluA virus-positive samples were tested. Of these, 104 had been subtyped previously; 54 were H3, 46 were 09H1, and 4 were fsH1. All were appropriately subtyped by both LDTs. Of the untyped FluA virus samples, three were subtyped as H3 by both LDTs and two were subtyped as H3 by the LDT-FAST assay only. The sample not subtyped by either LDT was retested with the FFABR assay and was now negative. Limit-of-detection (LOD) analyses were performed with five FluA virus strains. The LDT-FAST LODs were similar to the FFABR assay LODs, while the exWHO-FAST LODs were higher for two H3N2 strains, findings that were explained by analysis of primer/probe homology. In conclusion, either FluA virus typing assay would be a valuable complement to the Panther Fusion respiratory menu given the performance of these LDTs, the system's full automation, and the ability to split eluates for both IVD and LDT testing.
引用
- Stellrecht KA, Cimino JL, Maceira VP, et al. The Panther Fusion System with Open Access Functionality for Laboratory-Developed Tests for Influenza A Virus Subtyping. J Clin Microbiol. 2020. PMID: 32229600. doi:10.1128/JCM.00188-20
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。