文献速递临床免疫

预混链霉亲和素磁珠与生物素化捕获分子减轻两种罗氏免疫分析中的生物素干扰

Mitigating biotin interference in two Roche immunoassays by premixing biotinylated capturing molecules with streptavidin coated beads.
PubMed 文献 · Clin Chim Acta · 2020-02-19

本研究测试预混链霉亲和素包被磁珠与生物素化捕获抗体能否减轻罗氏C肽和人生长激素(hGH)夹心免疫分析中的生物素干扰。将试剂M与R1预混后,对生物素添加至1055 ng/mL的患者标本进行检测。原试剂中生物素使C肽结果降低88%、hGH降低97%;试剂改造后C肽干扰几乎消除,hGH仍降低约20%。

要点

  • 预混链霉亲和素磁珠与生物素化捕获分子可有效减轻某些免疫分析的生物素干扰。
  • 原试剂下1055 ng/mL生物素使C肽降低88%、hGH降低97%。
  • 试剂改造后C肽干扰几乎消除,hGH仍降低约20%。

核心结论

  • 预混链霉亲和素磁珠与生物素化捕获分子是减轻某些免疫分析生物素干扰的有效方法。

研究设计与样本

  • 对罗氏C肽和hGH夹心免疫分析试剂进行改造,将试剂M与R1预混后再分离。
  • 患者标本添加生物素至1055 ng/mL,用原试剂和改造试剂在罗氏cobas e411分析仪上检测。
  • 使用混合患者标本添加不同浓度生物素比较生物素干扰剂量反应曲线。

主要发现

  • 原试剂下1055 ng/mL生物素使C肽结果降低88%,hGH结果降低97%。
  • 试剂改造后该干扰效应对于C肽几乎消除,但hGH仍降低约20%。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Biotin is an interference in many streptavidin-biotin based immunoassays, causing falsely decreased results with sandwich immunoassays and falsely increased results with competitive immunoassays. It has been discussed that premixing streptavidin coated beads and biotinylated capturing molecules may prevent biotin interference. This study was designed to test whether such modification could mitigate biotin interference in two originally susceptible sandwich immunoassays. METHODS: Roche C-peptide and human growth hormone (hGH) immunoassays utilize three reagent containers for streptavidin coated beads (M), biotinylated capturing antibody (R1) and ruthenylated antibody (R2). The reagents were modified by premixing reagent M and R1. Following incubation, the beads were placed back in the M-container and R1-supernatant back to R1-container. Patient specimens were selected, spiked with biotin to 1055 ng/mL, and measured by both the original, unmodified reagent and modified reagent on Roche cobas e411 analyzer. The biotin interference dose response curves were also compared using pooled patient specimen spiked with different concentrations of biotin. RESULTS: For the original reagent, 1055 ng/mL of biotin decreased C- peptide results by 88% and hGH results by 97%. After reagent modification, this interference effect was nearly eliminated for C- peptide but remained about 20% decreased for hGH. CONCLUSION: Premixing streptavidin beads and biotinylated capturing molecules is an effective approach to mitigate biotin interference for certain immunoassays.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Yang J, Wiencek JR. Mitigating biotin interference in two Roche immunoassays by premixing biotinylated capturing molecules with streptavidin coated beads. Clin Chim Acta. 2020. PMID: 32084383. doi:10.1016/j.cca.2020.02.021

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。