多重流式免疫分析检测医护人员MMRV IgG的性能特征
评估BioPlex 2200 MMRV IgG多重流式免疫分析(MFI)与匹配的免疫荧光试验(IFA)在220名按ACIP标准判定为 presumptive 免疫的医护人员中的性能。结果显示Bio-Rad MFI对麻疹和腮腺炎IgG的敏感性显著低于Bion IFA,尤其在高疫苗接种率人群中。阴性结果可能导致不必要的再接种,并影响无法接种活疫苗者的就业。
要点
- Bio-Rad MFI对麻疹和腮腺炎IgG的敏感性显著低于Bion IFA。
- 在完全接种麻疹、腮腺炎、VZV疫苗的医护人员中,MFI阳性率分别为77.4%、84.8%、54.5%。
- 阴性结果可能误导再接种决策,影响无法接种活疫苗者的就业。
核心结论
- Bio-Rad MFI检测麻疹和腮腺炎IgG的敏感性显著低于Bion IFA,可能导致漏检 presumptive 免疫。
研究设计与样本
- 在220名按ACIP标准分类的医护人员中,比较BioPlex 2200 MMRV IgG MFI与匹配IFA的阳性率。
主要发现
- 在 presumptive 免疫者中,Bio-Rad MFI对麻疹、腮腺炎、风疹、VZV的阳性率分别为77.3%、85.4%、84.3%、91.1%。
- Bion IFA对麻疹、腮腺炎、VZV的阳性率分别为92.9%、91.1%、93.5%,无风疹IFA。
- 完全接种麻疹、腮腺炎、VZV者中,Bio-Rad MFI/Bion IFA阳性率分别为77.4%/93%、84.8%/90.7%、54.5%/90.9%。
- 完全接种风疹者中,Bio-Rad MFI阳性率为83.7%。
- 末次接种在入组前15年内者,Bio-Rad麻疹、腮腺炎、风疹IgG MFI阳性率分别为83.3%、93.3%、74.2%。
关键数值与推荐
- ACIP标准用于指导疫苗接种或抗MMRV IgG检测。
- Bio-Rad MFI对麻疹和腮腺炎IgG的敏感性显著降低。
临床怎么用
- 解读Bio-Rad MFI阴性结果时需谨慎,尤其对麻疹和腮腺炎,避免不必要的再接种。
- 对无法接种活疫苗者,MFI假阴性可能影响就业判定。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Immunity to measles, mumps, rubella, and varicella-zoster viruses (VZV; MMRV) is a common condition of employment for health care workers (HCWs) to ensure compliance with national standards and state laws. When documentation of complete vaccination or laboratory-confirmed infection is not available, Advisory Committee on Immunization Practices (ACIP) criteria are used to guide vaccination or anti-MMRV IgG testing. We assessed the performance of the BioPlex 2200 MMRV IgG multiplex flow immunoassay (MFI; Bio-Rad Laboratories, Hercules, CA) and matched immunofluorescence assays (IFAs; MBL Bion, Des Plaines, IL) in 220 HCWs categorized by ACIP criteria for presumptive immunity to MMRV. Among HCWs presumptively immune to measles, mumps, rubella, and VZV, the Bio-Rad MFI was positive in 77.3, 85.4, 84.3, and 91.1% of HCWs, respectively. Comparatively, the Bion IFA was positive in 92.9, 91.1, and 93.5% of HCWs presumptively immune to measles, mumps, and VZV (a rubella IFA was unavailable). Among HCWs fully vaccinated against measles, mumps, and VZV, Bio-Rad MFI/Bion IFA positivity rates were 77.4%/93%, 84.8%/90.7%, and 54.5%/90.9%, respectively. The Bio-Rad MFI was positive in 83.7% of HCWs fully vaccinated against rubella. For HCWs whose last vaccination event occurred within 15 years of enrollment, 83.3, 93.3, and 74.2% were positive by the Bio-Rad measles, mumps, and rubella IgG MFIs, respectively. We show significantly decreased Bio-Rad MFI sensitivity for detection of anti-measles and anti-mumps IgG-class antibodies in presumptively immune or fully vaccinated HCWs. Although negative results typically prompt revaccination, failure to recognize presumptive immunity in individuals unable to receive live, attenuated vaccines may have employment implications.
引用
- Theel ES, Sorenson M, Rahman C, et al. Performance Characteristics of a Multiplex Flow Immunoassay for Detection of IgG-Class Antibodies to Measles, Mumps, Rubella, and Varicella-Zoster Viruses in Presumptively Immune Health Care Workers. J Clin Microbiol. 2020. PMID: 32024728. doi:10.1128/JCM.00136-20