WASPLab分离软件自动分析尿培养的评估
该研究评估Copan公司开发的菌落分离软件在标准血琼脂和麦康凯琼脂上区分尿培养显著生长与无生长的性能。三个站点共12,931份尿培养经WASP接种、WASPLab培养并24小时成像,软件判读与实验室验证方案比较,总体敏感性和特异性分别为99.8%和72.0%。软件高敏感性可用于批量复核阴性培养以改善工作流程。
要点
- 分离软件对尿培养显著生长的敏感性达99.8%。
- 特异性为72.0%,阴性结果需结合专家规则复核。
- 软件可辅助批量复核阴性培养,优化实验室流程。
核心结论
- 图像分析软件区分尿培养显著生长与无生长的敏感性高,可用于批量复核阴性培养。
研究设计与样本
- 多站点评估Copan菌落分离软件,三个站点标本经WASP接种、WASPLab培养,0和24小时成像。
- 共12,931份尿培养,软件24小时图像菌落计数结果与实验室验证方案比较。
主要发现
- 总体敏感性和特异性分别为99.8%和72.0%。
- 二次复核显示91.1%的手工阳性/自动阴性标本因专家规则报告为污染或仅正常菌群,而非阈值计数。
- 9份手工阳性/自动阴性标本中,8份因软件编程忽略的微菌落生长,1份因菌落计数接近显著性限。
关键数值与推荐
- 敏感性99.8%,特异性72.0%(12,931份尿培养)。
临床怎么用
- 实验室可利用该软件批量复核阴性尿培养,提升工作效率。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Automation of the clinical microbiology laboratory has become more prominent as laboratories face higher specimen volumes and understaffing and are becoming more consolidated. One recent advancement is the use of digital image analysis to rapidly distinguish between chromogenic growth for screening bacterial cultures. In this study, colony segregation software developed by Copan (Brescia, Italy) was evaluated to distinguish between significant growth and no growth of urine cultures plated onto standard blood and MacConkey agars. Specimens from 3 sites were processed on a WASP instrument (Copan) and incubated on the WASPLab platform (Copan), and plates were imaged at 0 and 24 hours postinoculation. Images were read by technologists following validated laboratory protocols (VLPs), and results were recorded in the laboratory information systems (LIS). Image analysis performed colony counts on the 24-hour images, and results were compared with the VLP. A total of 12,931 urine cultures were tested and analyzed with an overall sensitivity and specificity of 99.8% and 72.0%, respectively. After secondary review, 91.1% of manual-positive/automation-negative specimens were due to expert rules that reported the plate as contaminated or growing only normal flora and not due to threshold counts. Nine specimens were found to be manual-positive/automation-negative; a secondary review demonstrated that the results of 8 of these specimens were due to growth of microcolonies that were programmed to be ignored by the software and 1 were due to a colony count near the limit of significance. Overall, the image analysis software proved to be highly sensitive and can be utilized by laboratories to batch-review negative cultures to improve laboratory workflow.
引用
- Faron ML, Buchan BW, Relich RF, et al. Evaluation of the WASPLab Segregation Software To Automatically Analyze Urine Cultures Using Routine Blood and MacConkey Agars. J Clin Microbiol. 2020. PMID: 31941690. doi:10.1128/JCM.01683-19
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。