NG-Test MCR-1侧流免疫分析与EDTA-黏菌素肉汤纸片洗脱法检测革兰阴性菌质粒介导黏菌素耐药的评估
本研究评估NG-Test MCR-1侧流免疫分析(LFA)和EDTA-黏菌素肉汤纸片洗脱(EDTA-CBDE)筛查法检测革兰阴性菌质粒介导黏菌素耐药(PMCR)。MCR-1 LFA阳性符合率100%、阴性符合率96.1%;EDTA-CBDE总体阳性符合率100%、阴性符合率94.3%。两种方法准确且易用。
要点
- MCR-1 LFA检测PMCR的阳性符合率100%,阴性符合率96.1%。
- EDTA-CBDE筛查法总体阳性符合率100%,阴性符合率94.3%。
- 两种方法均可用于公共卫生和临床微生物实验室检测PMCR。
核心结论
- MCR-1 LFA和EDTA-CBDE方法检测PMCR均准确且易用。
主要发现
- MCR-1 LFA检测PMCR的阳性符合率和阴性符合率分别为100%和96.1%。
- EDTA-CBDE筛查法总体阳性符合率和阴性符合率分别为100%和94.3%。
- EDTA-CBDE法在肠杆菌目中阴性符合率为94.2%,在铜绿假单胞菌中为96.0%。
关键数值与推荐
- MCR-1 LFA检测PMCR:阳性符合率100%,阴性符合率96.1%。
- EDTA-CBDE筛查法检测PMCR:总体阳性符合率100%,阴性符合率94.3%。
临床怎么用
- 可在公共卫生和临床微生物实验室实施MCR-1 LFA和EDTA-CBDE方法,以鉴别黏菌素耐药革兰阴性菌的耐药机制并检测PMCR用于感染预防控制。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Plasmid-mediated colistin resistance (PMCR) is a global public health concern, given its ease of transmissibility. The purpose of this study was to evaluate two methods for the detection of PMCR from bacterial colonies: (i) the NG-Test MCR-1 lateral flow immunoassay (LFA; NG Biotech, Guipry, France) and (ii) the EDTA-colistin broth disk elution (EDTA-CBDE) screening test method. These methods were evaluated using a cohort of contemporary, clinical Gram-negative bacillus isolates from 3 U.S. academic medical centers (126 isolates of the Enterobacterales, 50 Pseudomonas aeruginosa isolates, and 50 Acinetobacter species isolates; 1 isolate was mcr positive) and 12 mcr-positive CDC-FDA Antibiotic Resistance (AR) Isolate Bank isolates for which reference broth microdilution colistin susceptibility results were available. Eleven (4.6%) isolates were strongly positive by the MCR-1 LFA, with an additional 8 (3.4%) isolates yielding faintly positive results. The positive percent agreement (PPA) and negative percent agreement (NPA) for MCR-1 detection were 100% and 96.1%, respectively. Upon repeat testing, only a single false-positive MCR-2 producer remained, as the isolates with initially faintly positive results were negative. The EDTA-CBDE screening method had an overall PPA and NPA of 100% and 94.3%, respectively. The NPA for the EDTA-CBDE method was slightly lower at 94.2% with Enterobacterales, whereas it was 96.0% with P. aeruginosa The MCR-1 LFA and EDTA-CBDE methods are both accurate and user-friendly methods for the detection of PMCR. Despite the rarity of PMCR among clinical isolates in the United States, these methods are valuable tools that may be implemented in public health and clinical microbiology laboratories to further discern the mechanism of resistance among colistin-resistant Gram-negative isolates and to detect PMCR for infection prevention and control purposes.
引用
- Fenwick AJ, Bergman Y, Lewis S, et al. Evaluation of the NG-Test MCR-1 Lateral Flow Assay and EDTA-Colistin Broth Disk Elution Methods To Detect Plasmid-Mediated Colistin Resistance among Gram-Negative Bacterial Isolates. J Clin Microbiol. 2020. PMID: 31996440. doi:10.1128/JCM.01823-19