Hb Andrew-Minneapolis两例:HbA1c水平因检测方法而异
报告两例Hb Andrew-Minneapolis病例,均为β链144Lys→Asn杂合突变,糖耐量与糖化白蛋白正常。HPLC标准模式测得HbA1c分别为6.8%和6.5%,免疫法均为4.6%(正常低值)。
要点
- Hb Andrew-Minneapolis可致HPLC法HbA1c假性升高而免疫法呈正常低值。
- 两例糖耐量与糖化白蛋白均正常,提示HbA1c结果与方法相关。
- 发现HbA1c与血糖不符时,应考虑血红蛋白变异体干扰并更换方法验证。
核心结论
- Hb Andrew-Minneapolis病例中HPLC标准模式HbA1c升高而免疫法呈正常低值,可能与变异血红蛋白相关的修饰血红蛋白有关。
研究设计与样本
- 病例报告,共2例:1例25岁男性,1例32岁妊娠女性。
主要发现
- 两例糖耐量正常且糖化白蛋白在参考范围内。
- HPLC标准模式HbA1c分别为6.8%和6.5%,免疫法均为4.6%。
- 球蛋白基因分析检出杂合β链突变(β144Lys→Asn),诊断为Hb Andrew-Minneapolis。
- 病例1高分辨率HPLC色谱图显示多个异常峰,体外糖化反应后除HbX0和HbX1c外还见两个未知峰。
- 病例1校正修饰血红蛋白后免疫法HbA1c升至4.9%,仍在参考范围内。
关键数值与推荐
- HPLC标准模式HbA1c:病例1为6.8%,病例2为6.5%。
- 免疫法HbA1c:两例均为4.6%;病例1校正修饰血红蛋白后为4.9%。
- 基因分析:两例均为杂合β链突变β144Lys→Asn。
临床怎么用
- 当HbA1c结果与血糖或糖化白蛋白不一致时,应警惕血红蛋白变异体干扰。
- 对疑似变异体病例,可结合HPLC色谱图异常峰及免疫法结果综合判断。
局限性
- 未知峰的具体性质未鉴定。
- 仅2例病例报告,结论外推受限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
We experienced two cases of Hb Andrew-Minneapolis with high or low-normal HbA1c levels depending on the measurement method. Case 1 was a 25-year-old male, and case 2 was a 32-year-old pregnant woman. Both cases showed normal glucose tolerance levels and glycated albumin within the reference range. In both cases, the high-performance liquid chromatography (HPLC) method (standard mode) showed high HbA1c levels of 6.8% and 6.5%, respectively, while the HbA1c levels measured by immunoassay were low normal at 4.6% in both cases. Globin gene analysis detected heterozygous β-chain mutations (β144Lys → Asn) in both cases, which resulted in the diagnosis of Hb Andrew-Minneapolis. In case 1, a high-resolution HPLC chromatogram showed multiple abnormal peaks; two unknown peaks in addition to variant hemoglobin (HbX0) and glycation products of variant hemoglobin (HbX1c) were observed after in vitro glycation reaction. Although the details of unknown peaks were not identified, those might be modified hemoglobin associated with variant hemoglobin. The presence of unknown peaks could cause high HbA1c levels measured by HPLC (standard mode). Furthermore, the HbA1c level measured by immunoassay was increased to 4.9% within the reference range after adjustment for modified hemoglobin in case 1. Consequently, the high HbA1c levels measured by HPLC (standard mode) and the low-normal HbA1c level measured by immunoassay might be due to modified hemoglobin associated with variant hemoglobin.