文献速递临床免疫质量管理与危急值

Immulite® 2000XPi 总IgE与特异性IgE的第三方验证

Third-party verification of total and specific immunoglobulin E on analyzer Immulite® 2000XPi.
PubMed 文献 · Clin Chim Acta · 2019-10-10

该研究依据CLSI EP15-A3、EP9-A3和EP28-A3c指南,对Siemens Immulite®2000XPi过敏检测的分析性能进行第三方验证,包括稳定性、精密度、参考区间、真实度及总IgE两种方法可比性。结果显示大多数过敏原精密度符合厂家规格,总IgE样本2-8°C稳定7天,成人及儿童总IgE参考区间可采纳,但荧光与化学发光酶免疫法总IgE结果不可比。

要点

  • 总IgE样本在2-8°C可稳定7天,长于厂家声明。
  • 多数过敏原精密度符合厂家规格,总IgE和w1除外。
  • 荧光与化学发光酶免疫法总IgE结果不可比。

核心结论

  • Immulite®2000XPi检测的精密度和真实度可接受,总IgE样本稳定性优于厂家声明,成人及儿童参考区间可安全采纳,但总IgE两种方法结果不可比。

研究设计与样本

  • 依据CLSI EP15-A3、EP9-A3和EP28-A3c指南进行的第三方验证研究,使用血清混合样本评估稳定性、精密度、参考区间、真实度及方法可比性。

主要发现

  • 总IgE分析样本在2-8°C稳定7天。
  • 大多数过敏原精密度符合厂家规格,但总IgE(92.5 kIU/L,CVs 5.6%和7.6%)和w1(9.3 kU/L,CV=5.9%)不符合。
  • Bland-Altman图显示两种方法间总IgE存在统计学显著的恒定偏倚(-32.4 kIU/L)和比例偏倚(-20.2%)。
  • Passing Bablok显示两台Immulite分析仪间betula verrucosa(恒定和比例)、屋尘螨和蛋清(比例)偏倚有统计学意义但无临床意义。
  • 成人和儿童总IgE结果均在Siemens声明的参考区间内。

关键数值与推荐

  • 验收标准基于厂家规格和国家外部质量评估提供者数据。
  • 总IgE在92.5 kIU/L时CVs为5.6%和7.6%,w1在9.3 kU/L时CV为5.9%,超出厂家规格。
  • 两种方法间总IgE恒定偏倚-32.4 kIU/L,比例偏倚-20.2%。

临床怎么用

  • 实验室可采纳Siemens声明的总IgE参考区间用于成人和儿童。
  • 总IgE样本在2-8°C可保存7天。
  • 荧光与化学发光酶免疫法总IgE结果不可互换,需固定方法进行随访。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Although Immulite®2000XPi (Siemens) allergy testing is routinely used in laboratories worldwide; a well-designed comprehensive third-party verification of its analytical performance has not yet been published. Our aim was to verify stability, precision and reference intervals for total and some frequent specific IgEs, test trueness for several most common allergens and assess the comparability of fluorescent- and chemiluminescent-enzyme immunoassay for total IgE. METHODS: Verification was based on EP15-A3, EP9-A3 and EP28-A3c CLSI guidelines. Stability was tested on serum pools stored at 2-8 °C, during 7 days. Manufacturer's reference intervals were verified for children and adults. Acceptance criteria were based on manufacturer specifications and national External Quality Assessment provider data. RESULTS: Samples for total IgE analysis were stable at 2-8 °C for 7 days. Precision was within manufacturer's specifications for most of tested allergens, except for total IgE (92.5 kIU/L, CVs 5.6% and 7.6%) and w1 (9.3 kU/L, CV = 5.9%). Bland-Altman plot revealed statistically significant constant (-32.4 kIU/L) and proportional bias (-20.2%) for total IgE between two methods. Passing Bablok showed statistically but not clinically significant biases between two Immulite analyzers for betula verrucosa (constant and proportional), house dust mite and egg white allergens (proportional). For adults and children, total IgE results were within reference intervals declared by Siemens. CONCLUSIONS: Immulite®2000XPi assay has acceptable precision and trueness. Samples for total IgE are stable even longer than declared by the manufacturer. Declared reference intervals can safely be adopted for both adult and children population in our laboratory. Fluorescent- and chemiluminescent-enzyme immunoassay results of total IgE are not comparable.

来源:Clin Chim Acta,查看原文。版权归原作者所有。

引用

  1. Vuljanić D, Dukić L, Krleža JL, et al. Third-party verification of total and specific immunoglobulin E on analyzer Immulite® 2000XPi. Clin Chim Acta. 2019. PMID: 31606400. doi:10.1016/j.cca.2019.09.015

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。