三种化学发光法与液相色谱串联质谱检测血清醛固酮的比较

Comparison of three different chemiluminescence assays and a rapid liquid chromatography tandem mass spectrometry method for measuring serum aldosterone.
PubMed 文献 · Clin Chem Lab Med · 2019-01-01

该研究回顾性纳入2017年6-8月北京协和医院221例疑似原发性醛固酮增多症患者血清样本,用LC-MS/MS和三种化学发光免疫分析系统(DiaSorin® XL、iSYS、Auto Lumo A2000)测定血清醛固酮。结果显示LC-MS/MS测定值显著低于三种化学发光法(p<0.0001),Bland-Altman平均偏差百分比为-69.3%至-49.2%,提示不同方法结果不能互换。

要点

  • LC-MS/MS测定的血清醛固酮水平显著低于三种化学发光免疫分析法。
  • 三种化学发光法与LC-MS/MS的Spearman相关系数为0.745至0.950。
  • 不同方法间偏差较大,血清醛固酮结果不能互换使用。

核心结论

  • 化学发光免疫分析与LC-MS/MS测定血清醛固酮存在显著不一致,不同方法结果不能互换。

研究设计与样本

  • 回顾性纳入2017年6-8月北京协和医院221例疑似原发性醛固酮增多症患者血清样本。
  • 用LC-MS/MS和三种化学发光免疫分析系统(DiaSorin® XL、iSYS、Auto Lumo A2000)测定血清醛固酮。
  • 采用Passing-Bablok回归、相关矩阵和Bland-Altman图评价三种化学发光法之间的一致性。

主要发现

  • 四种方法实验室内精密度为2.1%至9.4%,但一种化学发光法在低血清醛固酮样本中变异系数超过20.0%。
  • LC-MS/MS测定的血清醛固酮水平显著低于三种化学发光法(p<0.0001)。
  • 四种方法Spearman相关系数为0.745至0.950(p<0.0001)。
  • Bland-Altman图显示三种化学发光法与LC-MS/MS的平均偏差百分比为-69.3%至-49.2%。
  • 重新校准后方法间相关性未改善,但LC-MS/MS与DiaSorin® XL/iSYS在医学决定水平的偏差及偏差百分比有所改善。

关键数值与推荐

  • 四种方法实验室内精密度为2.1%至9.4%,一种化学发光法在低血清醛固酮样本中变异系数超过20.0%。
  • LC-MS/MS测定值显著低于三种化学发光法(p<0.0001)。
  • 四种方法Spearman相关系数为0.745至0.950(p<0.0001)。
  • 三种化学发光法与LC-MS/MS的平均偏差百分比为-69.3%至-49.2%。

临床怎么用

  • 临床解读血清醛固酮结果时应考虑检测方法差异,不同化学发光法与LC-MS/MS结果不可互换。
  • 对于疑似原发性醛固酮增多症患者,应固定使用同一检测系统进行随访比较。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Background This study aimed to quantify and compare serum aldosterone (sALD) levels through three different chemiluminescence immunoassays (CLIAs) and liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis. Methods Serum samples from 221 patients with suspected primary aldosteronism (PA) were retrospectively included in this study conducted at the Peking Union Medical College Hospital from June to August in 2017. sALD levels were determined using the LC-MS/MS method and three different CLIA systems, viz., DiaSorin® XL, iSYS and Auto Lumo A2000. Pooled fresh serum samples were used for recalibration. Passing-Bablok regression analysis, correlation matrix, and Bland-Altman plots were used to evaluate the concurrence among ALD levels determined using the three CLIAs. Results Within-laboratory precision of the four assays ranged from 2.1% to 9.4%, except the coefficient variation (CV) of one of the CLIAs, which exceeded 20.0% for samples with low sALD levels. sALD levels determined using LC-MS/MS were significantly lower than those determined using the other three CLIAs (p < 0.0001). Spearman's correlation coefficient of the four assays ranged from 0.745 to 0.950 (p < 0.0001). The Bland-Altman plot showed that the average bias (%) for the three CLIAs and LC-MS/MS ranged from -69.3 to -49.2. After recalibration, this correlation did not improve among the assays. However, the bias and bias percentage at the medical decision level improved between LC-MS/MS and DiaSorin® XL/iSYS. Conclusions Significant inconsistencies between the results of CLIAs and LC-MS/MS indicate that different sALD measures cannot be used interchangeably.